SC-30
User’s Manual
57
The SC-30 may store a total of 100 final results in reverberation time mode.
7.4.2 View register
This option allows you to display on the screen the final results of the registers stored in the
SC-30 memory. To use this option, access VIEW REGISTER on the SC-30 menu.
The SC-30 allows you to select, using the keys
,
and
, the register you
want to display, and an index will appear on-screen showing all the registers stored in the
memory (register date and time of the beginning of the measurement process).
Information on the decay curve is not available.
7.5 Technical Specifications
7.5.1 Measurement Range
The decay curves are measured from the equivalent levels (
linear averaging)
with integration
time:
Integration time of the equivalent level
63
Hz
40 ms
125
Hz
20 ms
250
Hz
10 ms
500
Hz
10 ms
1
kHz
10 ms
2
kHz
10 ms
4
kHz
10 ms
The range of measurement for the T
20
and T
30
functions are:
T
20
63
Hz
0.48 s to 17.1 s
125
Hz
0.24 s to 17.1 s
250
Hz
0.12 s to 17.1 s
500
Hz
0.12 s to 17.1 s
1
kHz
0.12 s to 17.1 s
2
kHz
0.12 s to 17.1 s
4
kHz
0.12 s to 17.1 s