SC-30
User’s Manual
55
The SC-30 also measures the decay curves of the sound pressure level, but this only can be
displayed by recording the final result of the measurement and downloading these data to
the computer through the
CAPTURE Studio software application.
At the next graphic you can see all the Reverberation Time measurement process.
7.3.5 Stopping the measurement
While the measurement is in progress, it can be stopped by pressing
.
The state of
measurement indicator will switch from to .
7.3.6 Overload and Over range indicator
The
SC-30 has an overload indicator. While there is an overload, the sign
will appear in
the bottom right-hand corner of the screen.
If during the measurement the sign
^ appears, the SC-30 will have not validate the measured
values and it can not pass to the next measurement step.
If the overload occurs during the decay curve measurement, the measurement will be
aborted automatically.