Description
Possible reason
Troubleshooting
Display contrast is bad
Navigate to Settings -> Contrast and press the
OK button until the contrast is well.
Autoclave runs out of
water during sterilization
Lid gasket defective
Lid gasket insert wrong
Autoclave not locked correct
Steam release cock is open
…
Check if the gasket has been placed correctly
into the lid. The text “Place This Side Facing
Down” on the gasket must be readable for
the user. Check if the red knob on the handle
is in locking position during sterilization. For
cycles with sterilization time longer than 60
minutes, close the steam release cock.
Water is dropping out of
the gasket of the lid
On heat-up, the gasket of the lid will
be compressed. If the gasket is wet
also, this can cause a few drops
coming out of the gasket during heat-
up.
If the dropping doesn’t stop after heat-up,
please remove the gasket and clean it with
distilled water. If the problem is not resolved,
replace the lid gasket.
If the errors cannot be resolved contact your dealer or the manufacturer