![CeraGem CGM MB-1901 Manual Download Page 25](http://html1.mh-extra.com/html/ceragem/cgm-mb-1901/cgm-mb-1901_manual_2583446025.webp)
47
Guidance and Manufacturer’s declaration
- electromagnetic emission
The Model CGM MB-1901(CERAGEM CO., LTD. Automatic Thermal Massager)is
intended for use in the electromagnetic environment specified below.The
customer or the user of CGM MB-1901 should assure that it is used in such an
environment
Emission test
Compliance
Electromagnetic environment - guidance
RF emissions
CISPR 11
Group 1
The Model CGM MB-1901 uses RF energy
only for its internal function. Therefore. Its
RF emissions are very low and are not likely
to cause any interference in nearby electronic
equipment
Class B
Harmonics emission
IEC 61000-3-2
Voltage fluctuation
IEC 61000-3-3
A
Complies
The Model CGM MB-1901 is suitable for use
in all establishments, including domestic
establishments and those directly connected
to the public low-voltage power supply
network that supplies buildings used for
domestic purposes.
46
1. Explanation of mark, delivery, and storage
2. Attached documents
Classification
Form of electric shock protection
Degree of electric shock protection
Degree of protection from harmful invasion of water
Depending on the mode of operation
Continuous operation.
CLASS
II
EQUIPMENT
TYPE BF APPLIED PART.
Ordinary
Do not disassemble and contact the customer service center when problem occurs with a fuse.
* Pull the marked area to exchange the fuse (related to service engineer)
* The product has to be installed at place built protective device having appropriate breaking capacity so
that prospect short-circuit current do not exceed 40 A .
Capacity of a fuse : AC 250V, T4AL
Time-Lag,
4A,
Low breaking Capacity(Glass Body Fuse)
: Conformite Europeenne Mark.(Medical Devices)
CERAGEM
The disposal of waste products / residues, etc.
4
V~ : AC (alternating current) voltage source
ㅣ :
On (power: connectino to the mains)
O : Off (power: disconnection from the mains)
Authorised representative
in the european community
TYPE BF APPLIED PART
Serial number
Refer to instruction manual/booklet sign
General prohibition sign
Manufacturing date
Manufacturer
CLASS II EQUIPMENT
Fragile (handle with care)
Use no hooks
This way up
Keep dry
IP20
The product's IP Classification
When 'waterproof cover' cover to the product
Catalogue Number
Medical Device
IP21
Transport and
storage ambient
temperature range
60℃
-20℃
10%
95%
500hpa
1060hpa
Transport and
storage relative
humidity range
Transport and
storage atmospheric
pressure range
Unique device identifier
Model number
KR
Country of manufacture
English