background image

WHEN TO CHECK RUN-OUT:

Any time the probe is removed from its holder, the stylus is changed or the probe is used in a different 
machine, the alignment procedure should be repeated to ensure accuracy.  The user should also repeat 
the alignment procedure if the unit is dropped or receives any sudden external shock.  It is good practice
to periodically check alignment for quality control and to establish a base line maintenance schedule.

ANGULAR ALIGNMENT:

Angular alignment is the relation of the spindle center line to the stylus stem center line.  This 
adjustment provides for equal clearance all around the stylus stem.  This adjustment is rarely needed 
except when extremely small or long stylus is used.  When the stylus tip and stem are nearly the same 
diameter this adjustment is critical to keep the stem from contacting a vertical surface before the tip. 
This adjustment is also convenient to adjust parallelism when flat disc or block stylus types are used.    
The range of adjustment is very small as it is only intended to correct the stylus and spindle center line 
continuity.  If this adjustment must be performed use the run-out adjustment set-up shown above to 
monitor the effect of the adjustment.  The angular alignment and run-out adjustments are 
interdependent.  It will be necessary to repeat each several times to achieve significant results. When 
completed, the run-out at the stylus mount end of the stylus stem should be nearly the same as the run-
out at the tip.  Adjust the stylus tip run-out last.

To check the angular alignment begin by completing the “Dialing-in” procedures for the probe tip.  
Measure the run-out at the base of the stylus stem (the end away from the tip) with the dial indicator. If 
the run-out is greater than the difference between the tip diameter and stem diameter  then it will be 
possible for the stem to contact a vertical surface before the tip (also called “shanking out”). 

The angular alignment is accomplished by first performing a “Dialing-in” procedure on the stem base 
using the run-out adjustment set screws. Next, the tip is dialed-in using the angular adjustment set 
screws in the end cap of the probe.  These two steps are repeated until the difference in run-out between
the tip and stem base is acceptable. Be sure all set screws are equally tight when finished.

OUTPUT SWITCH LOGIC SELECTION:

The DP-4 output is a solid state switch that can be configured for normally open or closed logic. This 
choice can be specified when ordering (recommended) or can be changed in the field.  The standard 
configuration for the stock DP-4 is normally open.  A jumper block located under the mounting shank 
inside the probe sets the logic.  To access the jumper block unplug the probe, remove the stylus, loosen 
the run-out adjustment screws and remove the mounting shank.  Be careful not do damage any wires or
allow any dirt or debris inside the probe. Move the jumper from its present position to the opposite end 
of the three pin header to invert the logic.  Use tweezers or small pliers to slide the jumper block up off 
the pins and then slide it down on the pins at the opposite end of the three pin header.  Install the 
mounting shank and snug the run out adjustment screws.  The run out adjustment must be performed 
any time the mounting shank is removed.  

5

Summary of Contents for DP-4

Page 1: ...DP 4 Touch Probe Operator s Manual U S Patent 6553682 Rev 2019 05 28 Copyright 2003 2020 Centroid Corporation Howard PA 16841...

Page 2: ...nfirm machine is responding to DP 4 input by running an automatic cycle without the probe in the spindle and manually touching the DP 4 stylus mount boss Refer to control operator s manual on how to r...

Page 3: ...Diameter Calibration Page 6 Performance Characteristics Page 8 Maintenance and Care Page 9 Shipping Instructions Page 9 Kit Contents Page 10 Common Cables Page 11 Connection Diagram universal Page 12...

Page 4: ...orms OUTPUT FEATURES The output signal is a sink type compatible with most control inputs that can be configured for normally open or closed logic See CONNECTION DIAGRAM UNIVERSAL section This choice...

Page 5: ...2 56 4mm Mounting shank diameter and length D 0 5 12 7mm L 1 47 37 2mm Stylus mount thread M3 Output signal Switch FET open drain 30V 50mA Output logic Selectable normally open or closed Power supply...

Page 6: ...he DP 4 Probe into the dedicated 0 500 tool holder STEP 2 Install the complete assembly into the milling machine spindle without the retractile cord STEP 3 Install the 2 5 or 5mm stylus in your DP 4 p...

Page 7: ...then begin at the highest point of the run out and set the dial indicator so it is at the limit of its range at this point STEP 6 Rotate the probe so that the nearest run out adjustment set screw is...

Page 8: ...nt begin by completing the Dialing in procedures for the probe tip Measure the run out at the base of the stylus stem the end away from the tip with the dial indicator If the run out is greater than t...

Page 9: ...n the desired direction and finally Cycle Start The control will jog the stylus in the direction of the single axis move selected Touch the stylus once and the control will stop and reverse travel the...

Page 10: ...appear on the screen that will display the measured diameter of the bore Subtract this number from the known diameter of the gauge bore The resulting number will be the calibrated diameter of the prob...

Page 11: ...tion procedures then accuracy becomes dependent on repeatability PRE TRAVEL VARIATION RADIAL DIGITIZING PRECISION BORE Typical pre travel variation for DP 4 probe showing variation over 720 degrees of...

Page 12: ...d in or aligned then the probe should be sent in for evaluation and repair Probing errors may be caused by many factors If the errors are random in nature check surface condition of stylus tip and pro...

Page 13: ...auge p n 3193 Size and type may vary 5mm Ruby tip stylus M3 thread 40mmlength p n 3360 3 32 hex wrench p n 2898 DP 4 Probe complete kit with protective case and accessories Inset shows outside of case...

Page 14: ...are needed for new Centroid TM systems See the appropriate appendix in this manual for additional required adapters connection diagrams and parameter settings for your installation 11 Probe Cable P N...

Page 15: ...as follows 1 Power DC voltage supply positive terminal connected to pin 2 negative to pin 5 2 Signal Current sinking type pin 3 appears connected to pin 5 when signal is active Pin 3 appears unconnect...

Page 16: ...have a much better feel for where these parameters need to be set to achieve the results to meet your job requirements The X Y and Z stepover The smaller the stepover distance the longer the digitizin...

Page 17: ...end of the program You can also use the text editor to turn coolant on and off by adding M7 MIST or M8 FLOOD at the beginning of the program and M9 COOLANT OFF at the end Remember you can have only on...

Page 18: ...n enable 155 0 Probe type enable 186 1 Probe stuck retry disable 366 1 Probe Deceleration Multiplier Probe Jog Parameters control jog speed when probe is connected Probe must decelerate to stop when c...

Page 19: ...ate 15 3 76 2 Slow Probing Rate 16 5 127 Maximum Search Distance 18 50771 PLC input Probe Detect Spindle Inhibit 120 0 020 0 508 Probe stuck clearance amount 121 0 020 0 508 Grid digitizing minimum Z...

Page 20: ...unt nominal 14 30 762 Fast Probing Rate 15 3 76 2 Slow Probing Rate 16 5 127 Maximum Search Distance 18 15 PLC input Probe Detect Spindle Inhibit 120 0 020 0 508 Probe stuck clearance amount 121 0 020...

Page 21: ...ose that require a current source signal The circuit shown below can be used to convert the probe tripped signal from current sink to current source if necessary The resistor value is calculated to pr...

Reviews: