background image

                                                                                                                             

Data Logger  Instruction Manual 

 

 

6

 

bit3:

 1

Indicates the LCD is displaying time. 

bit4:

 no use

 

bit5:

 1

HOLD, 0

not HOLD 

bit6:

 1

LOW BATTERY , 0

BATTERY NORMAL 

bit7:

 1

°C 0

°F

 

3

th

  BYTE: 

bit7 

bit6 

bit5 bit4

bit3

bit2 bit1 

bit0 

Auto Power Off 

memory full

resolution

sign

OL

resolution  sign 

OL 

bit0:

 1

→T1 

is OL, 0

not OL 

bit1:

 1

→T1

 value is minus, 0

→T1

 value is plus. 

bit2:

 1

4

th

 byte and 5

th

 byte represent #### , 0

4

th

 byte and 5

th

 byte represent ###.# 

bit3:

 1

→T2 

is OL, 0

not OL 

bit4:

 1

→T2

 value is minus, 0

→T2

 value is plus. 

bit5:

 1

→8

th

 byte and 9

th

 byte represent #### , 0

→8

th

 byte and 9

th

 byte represent ###.# 

bit6:

 1

Memory is full. 0

Memory is not full. 

bit7: 

1

Auto power off enabled. 0

Auto power off disabled.           

4

th

  BYTE: 

first two BCD code of T1 value.

 

5

th

  BYTE:

last two BCD code of T1 value 

6

th

  BYTE:  

  If bit3 of 

2

nd

 BYTE

 =0 : first two BCD code of T1-T2 value. 

  If bit3 of 

2

nd

 BYTE

 =1 : two BCD code of month.

 

7

th

  BYTE: 

   

  If bit3 of 

2

nd

 BYTE

 =0 : last two BCD code of T1-T2 value. 

  If bit3 of 

2

nd

 BYTE

 =1 : two BCD code of day.

 

8

th

  BYTE: 

   

  If bit3 of 

2

nd

 BYTE

 =0 : first two BCD code of T2 value. 

  If bit3 of 

2

nd

 BYTE

 =1 : two BCD code of hour.

 

9

th

  BYTE: 

   

  If bit3 of 

2

nd

 BYTE

 =0 : last two BCD code of T2 value. 

  If bit3 of 

2

nd

 BYTE

 =1 : two BCD code of minute.

 

10

th

    BYTE:

 end byte, it value is 3, 1

nd 

and 10

th

 are used to check frame error. 

 

Appendix: Thermo couple probe specification   

Model Range 

Tolerances 

Description 

TP-K01 

Bead probe 

-50

 to 200

 

-58

 to 392

 

±

2.2

 or 

±

0.75

 

(

±

3.6

 or 

±

0.75

with Teflon tape insulation Maximum 
insulating temperature : 260

 

 

                                                         

 

TP-K01: 

probe for general condition measurements, especially for 

complex and hard to reach places.

 

Summary of Contents for 306

Page 1: ...T1 T2 REC POWER UP OPTIONS RANGE 200 C HOLD TIME SETUP CLOCK REC INTV C F 328 F 1370 C 2498 F 306 Datalogger SE 305 ...

Page 2: ...tion the Thermocouples 3 4 3 Selecting the Temperature Scale 3 4 4 Data Hold Operation 3 4 5 DataLogger 3 4 6 Clock Setup 3 4 7 Recording Interval Setup 4 4 8 Time Operation 4 4 9 MAX MIN Operation 4 4 10 Auto Power Off 4 4 11 Low Battery Condition 4 4 12 Calibration Point 4 4 13 Digital Output 5 V Setup SE305 Thermo DataLogger RS232 interface software 7 ...

Page 3: ...eading 1 C 200 C 400 C 0 5 reading 1 C 400 C 1370 C 0 2 reading 1 C 328 F 200 F 0 5 reading 2 F 200 F 200 F 0 2 reading 2 F 200 F 2498 F 0 3 reading 2 F Note The basic accuracy Specification does not include the error of the probe Please refer to the probe accuracy specification for additional details Sample Rate 1 25 times per second Dimension 184 64 30mm Weight 210g Approx Accessory K Type Bead ...

Page 4: ... year is displayed in the main window The Battery is not sufficient for proper operation REC This indicates that the tester is recording If it blinks it indicates the memory is full Button Location 1 K type temperature sensor T1 input connector 2 K type temperature sensor T2 input LCD display 3 ON OFF button 4 Time display button 5 Record button 6 MAX MIN function control button 7 HOLD button 8 C ...

Page 5: ... is no longer needed one may release the data hold operation by pressing HOLD button again When the meter is under Data Hold operation the TIME MAX MIN and C F button are disabled when you press TIME C F and max min button in HOLD mode there will be two continuous beeps To exit the MAX MIN mode one may press and hold MAX MIN button for two seconds 4 5 DataLogger When one presses the REC button the...

Page 6: ...ain MAX and MIN will blink together This means that all these data is updated in the memory and the reading is the present temperature One may press MAX MIN to circulate the display mode among these options When the meter is under MAX MIN operation and C F button are disabled when you press C F button in MAX MIN mode there will be two continuous beep To exit the MAX MIN mode one may press and hold...

Page 7: ...will return 3 0 6 ASCII 13 Command U Return 32768 bytes Command P Instead of returning all 32768 bytes it only return recorded data Command H Equivalent to one pushing on the HOLD button and no message is returned Command M Equivalent to one pushing on the MAX MIN button and no message is returned Command N Equivalent to one pushing and hold the MAX MIN button for two seconds to exit MAX MIN mode ...

Page 8: ...E first two BCD code of T1 value 5th BYTE last two BCD code of T1 value 6th BYTE If bit3 of 2nd BYTE 0 first two BCD code of T1 T2 value If bit3 of 2nd BYTE 1 two BCD code of month 7th BYTE If bit3 of 2nd BYTE 0 last two BCD code of T1 T2 value If bit3 of 2nd BYTE 1 two BCD code of day 8th BYTE If bit3 of 2nd BYTE 0 first two BCD code of T2 value If bit3 of 2nd BYTE 1 two BCD code of hour 9th BYTE...

Page 9: ...rt the SE305 CD ROM into your CD drive The SE305 installer should start automatically If it does not you can start it by running SETUP EXE from the root drive of the CD ROM 3 When installation is complete it will copy SE305 exe executable file and help file to your hard disk default is c program files SE305 4 For other operation instruction please refer to the on line help while executing SE305 Ma...

Page 10: ... present data Real Time Run Start collecting real time data Stop Stop collecting real time data DataLogger By opening the DataLogger Window the user can load recorded data of meter to PC in this window ComPort Select the port manually Option Range Change the Y axis extention Graph Customization Graph Customization For more operation instruction please refer to the online help while executing SE305...