Data Logger  Instruction Manual 

 

 

5

TX

RX

GND

 

4.13 Digital Output: 

The Digital Output is a 9600bps N 81 serial interface. 
The RX is a 5V normal high input port. 
The TX is a 5V normal high output port. 
 

The command of Digital Output is list below: 

RS232 command 

Function 

Remarks 

K(ASC 4BH) 

Ask for model No. 

Return 4 bytes 

A(ASC 41H) 

Inquire all encoded data 

Return encoded 10 byte 

H(ASC 48H) 

Hold button 

 

M(ASC 4DH) 

MAX/MIN button 

 

N(ASC 4EH) 

Exit MAX/MIN mode 

 

T(ASC 52H) 

TIME button 

 

C(ASC 43H) 

C/F button 

 

U(ASC 55H) 

Dump all memory of thermometer 

return 32768 bytes 

P(ASC 50H) 

Load recorded data 

 

Command K:   

Return 4 bytes. For example, when sending command "K" to the meter, it will return "3","0","6", 
ASCII(13) . 

Command U:   

Return 32768 bytes . 

Command P:   

Instead of returning all 32768 bytes, it only return recorded data . 

Command H:   

Equivalent to one pushing on the HOLD button and no message is returned. 

Command M:   

Equivalent to one pushing on the MAX/MIN button and no message is returned. 

Command N:   

Equivalent to one pushing and hold the MAX/MIN button for two seconds to exit MAX/MIN mode. 

Command T:   

Equivalent to one pushing on the TIME button and no message is returned. 

Command C:   

Equivalent to one pushing on the °C/°F button and no message is returned. 

Command A:

 

1

nd

 BYTE: 

The first byte is the start byte , it value is 2. 

2

nd

 BYTE: 

bit7 bit6  bit5 

bit4 

bit3 bit2 bit1

bit0 

C/F Low 

Bat Hold

  TIME  MAX/MIN  REC 

bit 0:

 1

recording mode, 0

not recording 

bit 2  bit 1  

  0          0               

normal mode 

  0          1               

MAXIMUM mode 

  1          0               

MINIMUM mode 

  1          1               

calculate MAX/MIN in background mode . 

Summary of Contents for 306

Page 1: ...T1 T2 REC POWER UP OPTIONS RANGE 200 C HOLD TIME SETUP CLOCK REC INTV C F 328 F 1370 C 2498 F 306 Datalogger SE 305 ...

Page 2: ...tion the Thermocouples 3 4 3 Selecting the Temperature Scale 3 4 4 Data Hold Operation 3 4 5 DataLogger 3 4 6 Clock Setup 3 4 7 Recording Interval Setup 4 4 8 Time Operation 4 4 9 MAX MIN Operation 4 4 10 Auto Power Off 4 4 11 Low Battery Condition 4 4 12 Calibration Point 4 4 13 Digital Output 5 V Setup SE305 Thermo DataLogger RS232 interface software 7 ...

Page 3: ...eading 1 C 200 C 400 C 0 5 reading 1 C 400 C 1370 C 0 2 reading 1 C 328 F 200 F 0 5 reading 2 F 200 F 200 F 0 2 reading 2 F 200 F 2498 F 0 3 reading 2 F Note The basic accuracy Specification does not include the error of the probe Please refer to the probe accuracy specification for additional details Sample Rate 1 25 times per second Dimension 184 64 30mm Weight 210g Approx Accessory K Type Bead ...

Page 4: ... year is displayed in the main window The Battery is not sufficient for proper operation REC This indicates that the tester is recording If it blinks it indicates the memory is full Button Location 1 K type temperature sensor T1 input connector 2 K type temperature sensor T2 input LCD display 3 ON OFF button 4 Time display button 5 Record button 6 MAX MIN function control button 7 HOLD button 8 C ...

Page 5: ... is no longer needed one may release the data hold operation by pressing HOLD button again When the meter is under Data Hold operation the TIME MAX MIN and C F button are disabled when you press TIME C F and max min button in HOLD mode there will be two continuous beeps To exit the MAX MIN mode one may press and hold MAX MIN button for two seconds 4 5 DataLogger When one presses the REC button the...

Page 6: ...ain MAX and MIN will blink together This means that all these data is updated in the memory and the reading is the present temperature One may press MAX MIN to circulate the display mode among these options When the meter is under MAX MIN operation and C F button are disabled when you press C F button in MAX MIN mode there will be two continuous beep To exit the MAX MIN mode one may press and hold...

Page 7: ...will return 3 0 6 ASCII 13 Command U Return 32768 bytes Command P Instead of returning all 32768 bytes it only return recorded data Command H Equivalent to one pushing on the HOLD button and no message is returned Command M Equivalent to one pushing on the MAX MIN button and no message is returned Command N Equivalent to one pushing and hold the MAX MIN button for two seconds to exit MAX MIN mode ...

Page 8: ...E first two BCD code of T1 value 5th BYTE last two BCD code of T1 value 6th BYTE If bit3 of 2nd BYTE 0 first two BCD code of T1 T2 value If bit3 of 2nd BYTE 1 two BCD code of month 7th BYTE If bit3 of 2nd BYTE 0 last two BCD code of T1 T2 value If bit3 of 2nd BYTE 1 two BCD code of day 8th BYTE If bit3 of 2nd BYTE 0 first two BCD code of T2 value If bit3 of 2nd BYTE 1 two BCD code of hour 9th BYTE...

Page 9: ...rt the SE305 CD ROM into your CD drive The SE305 installer should start automatically If it does not you can start it by running SETUP EXE from the root drive of the CD ROM 3 When installation is complete it will copy SE305 exe executable file and help file to your hard disk default is c program files SE305 4 For other operation instruction please refer to the on line help while executing SE305 Ma...

Page 10: ... present data Real Time Run Start collecting real time data Stop Stop collecting real time data DataLogger By opening the DataLogger Window the user can load recorded data of meter to PC in this window ComPort Select the port manually Option Range Change the Y axis extention Graph Customization Graph Customization For more operation instruction please refer to the online help while executing SE305...

Page 11: ...4 F NO 415 Jung Jeng Rd 238 Shu Lin Chien Taipei Taiwan E Mail center centertek com http www centertek com 306 03 CENTER TECHNOLOGY CORP ...

Reviews: