Edit References
: Need High, Middle, Low reference values to create Linear Fat Equation
.
Example:
12.20%, 10.56%, 5.30% Fat
Press
“
ADD
” for each reference value to be entered:
1. 12.20
2. 10.56
3. 5.30
Collect Standards Data: Choose item number “
1
”
to collect data for 12.20.
EDIT Standard - 12.20 (
Start with highest fat sample)
Press
“
ADD Standard
” to collect standards data for 12.20 to generate
signal/mass.
Using the12.20%F sample, repeat “Add Standard” step a minimum of 3 times (up
to 8).
Edit Standard – 12.20%
%Fat
S/M
Mass
------------
-----
------
*
1. -------
1539.6340
2.2315 g
*
2. -------
1534.0634
2.3152 g
*
3. -------
1537.0634
2.5315 g
Note:
Fat calculation will not be displayed until standards data has been collected for at
least two reference points (12.20 and 5.30).
After enough data points have been collected for 12.20, Press “
Prev Page
”
and choose next item number “
3
”
to collect data for the lowest fat sample (5.30).
EDIT Standard – 5.30 (lowest fat sample)
Press
“
ADD Standard
” and repeat at least 3 times to collect signal/mass.
Now that you have the second point, the fat values will be calculated and shown.
Select the points (
press line number to toggle on/off
) that are within specifications
of reference to create your Linear Fat Equation.
Note:
Each time you select a point it will recalculate the line and therefore recalculate
the other standards data (even those under 12.20).
Edit Standard – 5.30%
%Fat
S/M
Mass
------------
-----
----------
*
1. 5.26
1036.0111
2.1222 g
*
2. 5.35
1040.1713
2.2231 g
*
3. 5.30
1038.0225
2.1315 g
After enough data points have been collected, Press “
Prev Page
” and choose
next item number “
2
”
to collect data for the middle point (10.60).
Edit Standard – 10.56 (middle point to validate line)
Press
“
ADD Standard
” and repeat at least 3 times to collect signal/mass.
Select points (
press line number to toggle on/off
) that are within specifications.
Edit Standard – 10.56%
%Fat
S/M
Mass
------------
-----
----------
*
1. 10.56
1414.2478
2.4311 g
*
2. 10.48
1409.0634
2.3315 g
*
3. 10.61
1417.2236
2.5314 g
Summary of Contents for Smart Trac II
Page 1: ...Corporation Quick Start Guide 600269 Rev 1 ...
Page 8: ...System Setup ...
Page 9: ...SMART Trac II Setup Guide ...
Page 12: ...3 Right click SMARTNMR and select Tools System Setup ...
Page 13: ...4 Select on System Setup ...
Page 16: ...SHIM Setup prompt for Oil Standard Cal lock prompt for Normalization Standard ...
Page 23: ...Method Importation ...
Page 25: ...Method Creation ...