13 14
The displayed value is the difference between the reference value
and the measured value.
1. Perform the measurement as described in the operating
instructions.
2. Press the REL button to store the reading in the display and
the "REL" indicator will appear on the display.
3.The display will now indicate the difference between the stored
value and the measured value.
4.Press the REL button to exit the relative mode.
Meter Functions - Voltage (V)
The meter will automatically
select he best voltage (V) range.
Select DCV or ACV with the
MODE
button.
Insert:
Black lead in
COM
terminal.
Red lead in
V-
Ω
-PRM
terminal
Touch the Black probe to ground or
to the negative (-) circuit
Touch the Red probe to the circuit
coming from the power source
IMPORTANT: Voltage must be
measuring in parallel (Red probe
measuring circuit from power
source).
Accuracy
Selection of a lower range will
move the decimal point one
place and increase the accuracy.
An “OL” display means the
range is too low. Select the next
higher range.
Analog Bar Graph
The Bar Graph is easiest to
read when the data causes
the digital display to rapidly
change. It is also useful for
trend setting or directional
data.
WARNING
When measuring voltage,
be sure the Red test lead is
in the terminal marked “V”.
If the test lead is in an Amp
(A) or
Milliampere (mA) terminal,
you may be injured or the
meter damaged.