2
50V (25V) rms AC. The values in brackets are valid
forlimited ranges (as for example medicine and
agriculture).
Prior to measurement ensure that the test leads
and the test instrument are in perfect condition.
When using this instrument only the handles of
the probes may be touched – do not touch the
probe tips.
This instrument may only be used within the
ranges specified and within low voltage systems up
to 690V.
Prior to usage ensure perfect instrument function
(e.g. on known voltage source).
The voltage testers may no longer be used if one or
several functions fail or if no functionality is
indicated.
Do not use this instrument under damp conditions.
Perfect display I sonly guaranteed within a
temperature range of -10
℃
up to +55
℃
, at relative
humidity question <85%.
If the operator’s safety cannot be guaranteed, the
instrument must be removed from service and