• Press up/down button to view each data
• Press right button to return or press left button to delete this group
data.
•
5. Calibration and Measurement
The gauge can calibrate Zero-point of Fe or Non-Fe, or delete wrong
calibrated zero-point of Fe or Non-Fe.
It suggested that zero-point be calibrated before use.
6. Zero-point calibration of Fe or Non-Fe
•
Press left button to enter the menu
•
Press up/down button to select calibration
•
Press left button to enter view
•
Press up/down button to select zero-point calibration of Fe or Non-Fe
•
Press left button to enter
•
Press right button to return the measurement interface
•
Measure directly on Fe or Non-Fe matrix(can measure many times,
last measure is effective)
•
•
Press left button to confirm the calibration
Also you can press and hold the down button to enter the Zero-point
calibration of Fe in measuring mode
Also you can press and hold the up button to enter the Zero-point
calibration of Non-Fe in measuring mode
7. Delete Fe or Non-Fe Zero-point
•
Press left button to enter the menu
•
Press up/down button to select calibration
•
Press left button to enter calibration
•
Press up/down button to delete zero-point of Fe or Non-Fe
•
Press left button for Ok
•
Press right button to return
12
Coating Thickness Tester User Manual