background image

9

EN

NewTom GO 2D/3D 

- TECHNICAL MANUAL

EN

PC CONFIGURATION

3.3  OPERATING SYSTEM SETTINGS

Open the Control Panel, select “User Account” and 

then “Change User Account Control Settings”.

 

Select the lowest notification level “Never Notify”.

RESTART THE PC.

Summary of Contents for NewTom GO 2D

Page 1: ...Cod 97071219 Rev 01 01 2017 NewTom GO 2D 3D technical manual EN...

Page 2: ......

Page 3: ...SERVER SOFTWARE INSTALLATION 19 3 9 INTEL PRO 1000 DRIVER UPDATE 21 3 10 ACQUISITION SERVER CONFIGURATION 23 3 11 NNT SOFTWARE INSTALLATION AND CONFIGURATION 27 4 CALIBRATION See corresponding 3D CALI...

Page 4: ...2 EN NewTom GO 2D 3D TECHNICAL MANUAL CONTENTS...

Page 5: ...following 1 If the device is in AIR version check that the screws fastening the column to the wall are not too tightened and the feet are adjusted so that the column is levelled and is not affected by...

Page 6: ...LED only if using button 96609090 white wire 4 Emission LED only if using button 96609090 blue wire K14 External warning lights clean contacts no voltage provided Optional Use lamps up to 240 VAC max...

Page 7: ...OVAL 2 COVER HANDLES MIRROR REMOVAL 2 1 MONOBLOC COVER Remove the highlighted screws and pull the interlo cking cover 2 2 PANEL COVER Remove the highlighted screws and pull the interlo cking rear cove...

Page 8: ...OVER HANDLES MIRROR REMOVAL 2 3 MIRROR Slightly press the mirror against the column move the mirror upwards and turn it outwards to remove it 2 4 HANDLES Remove the accessory drawer and loosen the hi...

Page 9: ...ighlighted screws and remove the cover 2 6 SIDE COVERS Loosen the highlighted screws and remove the interlo cking side covers 2 7 ARCH COVER NOTE Before removing the arch cover it is necessary to remo...

Page 10: ...with the specifications and install the entire software autonomously The 2 PC network cards will be needed for the communication between device and PC and for the connection between Sensor and PC The...

Page 11: ...TECHNICAL MANUAL EN PC CONFIGURATION 3 3 OPERATING SYSTEM SETTINGS Open the Control Panel select User Account and then Change User Account Control Settings Select the lowest notification level Never N...

Page 12: ...NFIGURATION From the Control Panel select Power Options On the Balanced profile select Change plan settings Set Turn OFF display and Put the computer to sleep to Never Select Change advanced power set...

Page 13: ...h the mouse on any free area of the desktop with no icons or windows and select Personalize from the context menu Press on the Screen Saver icon in the bottom right corner and select None in the next...

Page 14: ...cal area network or directly to the Acquisition Server PC In both cases the 3D panel Ethernet connection must absolutely be point to point and is always managed autonomously by the software and the ma...

Page 15: ...d and press enter The connection management window will open If a fixed IP address must be set for the network card dedicated to communication right click on the corresponding connection not the INTEL...

Page 16: ...between 1 and 254 except for the device ip address 160 in case of default address The Subnet mask is already filled out properly Click OK to confirm On the other hand in case of connection to a local...

Page 17: ...ach panel is equipped with a specific installation CD containing data that are not interchangeable with other units Carefully keep the CD and make sure that at least one backup copy is available Inser...

Page 18: ...ra Driver are already present and will uninstall them if necessary WARNING this procedure may take a long time do not interrupt the process At the end the PC will restart and resume installation If th...

Page 19: ...xe file the screen with the list of network cards available in the PC will appear Only near Intel Pro 1000 board connected to the 3D panel it is necessary that in Current Driver column the eBUS Univer...

Page 20: ...k on Computer select Properties and select Device Manager in the left column Expand the Network adapters branch and right click on Intel PRO 1000 GT Desktop Adapters select Properties Select Advanced...

Page 21: ...t displayed directly in this window If it is not displayed search for the Performance Options item and click on Properties The Receive Buffers value is displayed there 3 8 ACQUISITION SERVER SOFTWARE...

Page 22: ...20 EN NewTom GO 2D 3D TECHNICAL MANUAL EN PC CONFIGURATION Select I accept the agreement then click Next Select the installation for 2D and 3D devices and click on Next Click on Install...

Page 23: ...000 DRIVER UPDATE The Intel PRO 1000 ethernet card is automatically recognized from the Operating System However the default system driver should be updated to a specific version in order to ensure th...

Page 24: ...22 EN NewTom GO 2D 3D TECHNICAL MANUAL EN PC CONFIGURATION Choose I accept and click Next Press Next and proceed with the default options Click on Install...

Page 25: ...t Click on Finish Reboot the PC 3 10 ACQUISITION SERVER CONFIGURATION Once the PC IP addresses have been properly set the Acquisition Server can be configured Clicking on the Acquisition Server icon i...

Page 26: ...concerning machine errors the second one bigger Clicking on the Settings icon at top right the connection and software option panel will open Select Change Mode and confirm the application restart mes...

Page 27: ...URATION Go back to Settings select Find Device The device will be searched automatically select it and confirm Select Connect Device and tick off the Auto Connect Device option Click on Close After a...

Page 28: ...C CONFIGURATION Now set the connection with the 3D panel Click again on Settings and on Find Sensor The device 3D panel will be searched automatically Select it and confirm Confirm by Close The acquis...

Page 29: ...o receive images from the 3D device is NNT version 6 5 or higher installed in its MAIN WORKSTATION version NNT must be installed on the Acquisition Server PC Start NNT installation select 3D device in...

Page 30: ...GO 2D 3D TECHNICAL MANUAL EN PC CONFIGURATION Wait and click on Install At the end restart the computer Upon first NNT launch the display mode has to be selected Oral and Maxillofacial Surgery is the...

Page 31: ...mode by pressing CTRL ALT SHIFT END simultaneously the Service Level message under NNT in the main screen confirms proper execution Press ALT F10 and enter the 336699 password Click on Browse in the w...

Page 32: ...ce Mode press ALT F10 and enter the 12345 password The window displayed in the figure will open Click on Find Select the device and click on Add selected Make sure that the settings are correct as dis...

Page 33: ...C CONFIGURATION After saving the settings the following NNT display will appear Click on the button in the top right corner and select the virtual keyboard from the menu Upon first launch the virtual...

Page 34: ...erver software and select Clients Allowed The list of all the clients that have attempted to connect to the virtual keyboard will be displayed Tick off the necessary one s and close Now restarting the...

Page 35: ...33 EN NewTom GO 2D 3D TECHNICAL MANUAL EN CONTROL BOARDS DESCRIPTION 5 CONTROL BOARDS DESCRIPTION 5 1 UNIT BLOCK DIAGRAM...

Page 36: ...NG The K17 input terminal board is always powered live as marked on the info plate below Before handling always cut the power via the external circuit breaker Once the power to the device has been cut...

Page 37: ...35 EN NewTom GO 2D 3D TECHNICAL MANUAL EN CONTROL BOARDS DESCRIPTION...

Page 38: ...v 6 3x32 230V F2 Standby power fuse FUSE 2AT 230V F1 48V output fuse FUSE 2 5AF 450V DC DIP Switches SW1 Always off reserved use 5 2 3 LEDS LIST LED ID COLOR NAME ON BOARD DESCRIPTION DL1 YELLOW ACT C...

Page 39: ...RDYL External ready lamp relay status Normally off on when unit is ready DL24 YELLOW XRYL External x ray on lamp relay status Normally off on during x ray emission DL25 GREEN BUT Xray button pressed...

Page 40: ...38 EN NewTom GO 2D 3D TECHNICAL MANUAL EN CONTROL BOARDS DESCRIPTION PIN 4 signal X ray ON...

Page 41: ...39 EN NewTom GO 2D 3D TECHNICAL MANUAL EN CONTROL BOARDS DESCRIPTION...

Page 42: ...witches DESCRIPTION SW4 Not used all OFF SW2 Reboot button SW3 Factory reset switch 5 3 3 LEDS LIST LED ID COLOR NAME ON BOARD DESCRIPTION DL1 YELLOW CAN ACT CAN activity Normally on flashing On stead...

Page 43: ...3 GREEN 24V 24V tension present normally on DL24 GREEN 5 0V 5V CPU tension present normally on DL25 GREEN 1 2V 1 2V CPU tension present normally on DL32 GREEN 48V R axis Motor tension present normally...

Page 44: ...42 EN NewTom GO 2D 3D TECHNICAL MANUAL EN CONTROL BOARDS DESCRIPTION 5 4 SENSOR INTERFACE BOARD 97661485...

Page 45: ...43 EN NewTom GO 2D 3D TECHNICAL MANUAL EN CONTROL BOARDS DESCRIPTION...

Page 46: ...5 4 1 CONNECTORS LIST CONNECTOR DESCRIPTION K1 Sensor synchronisation supply from LOGIC K2 24V DC Panel supply K3 Sensor synchronisation 5 4 2 LEDS LIST LED ID DESCRIPTION DL1 VIN 48V from LOGIC boar...

Page 47: ...45 EN NewTom GO 2D 3D TECHNICAL MANUAL EN CONTROL BOARDS DESCRIPTION 5 5 MONOBLOC AREA...

Page 48: ...EN CONTROL BOARDS DESCRIPTION 5 6 KEYBOARD BOARD 97661484 5 6 1 CONNECTORS LIST CONNECTOR DESCRIPTION K1 Keyboard signals from LOGIC 5 7 LOGO BOARD 97661539 5 7 1 CONNECTORS LIST CONNECTOR DESCRIPTIO...

Page 49: ...e centred with respect to the sagittal laser perform a mechanical adjustment Proceed as follows 1 Check that craniostat is centred 2 Close the 2 side tabs 3 Remove lower cover There are no screws just...

Page 50: ...l out and distribute the Report of assembly using the Form FDA 2579 upon completion of an installation and within 15 days of the date of installation For more guidance please refer to FDA document num...

Page 51: ......

Page 52: ......

Reviews: