10 APPENDIX A- Technical references
DLP table declared for operative modalities of the device
Values expressed in mGy
x
cm. Tolerance is
30%. Values calculated according to IEC 60601-2-44: 2009 Par 201.3.214 c)
FOV
Eco Scan
Standard Dose
Regular Scan
Standard Dose
Regular Scan
Boosted Dose
Enhanced Scan
Standard Dose
Enhanced Scan
Boosted Dose
[18x16]
48,5
72,2
145,8
96,9
192,2
[15x22e]
101,2
154,0
239,8
-
-
[15x12]
37,0
54,0
110,5
74,0
146,3
[12x8]
27,1
38,9
66,1
54,2
87,9
[8x8]
23,0
35,5
56,9
47,4
75,7
[15x5] HiRes
64,8
98,2
-
132,5
-
[12x8] HiRes
80,2
120,9
-
163,3
-
[8x8] HiRes
66,5
100,2
-
135,2
-
[6x6] HiRes
51,0
62,3
-
103,3
-
Date: 2013-07-11
34
Measured using “head phantom” according to IEC 60601-2-44: 2009 Par. 203.108
35
Loading factors: automatically set by software
Eco Scan Std Dose: 110kV 17.76mAs - Reg. Scan Std Dose: 110kV 26.35mAs - Reg. Scan Boosted Dose: 110kV 67.20mAs
- Enh. Scan Std Dose: 110kV 35.52mAs - Enh. Scan Boosted Dose: 110kV 87.64mAs
36
Loading factors: automatically set by software
Eco Scan Std Dose: 110kV 32,85mAs - Reg. Scan Std Dose: 110kV 48,84mAs - Reg. Scan Boosted Dose: 110kV 97,63mAs
37
Loading factors: automatically set by software
Eco Scan Std Dose: 110kV 19.07mAs - Reg. Scan Std Dose: 110kV 28.37mAs - Reg. Scan Boosted Dose: 110kV 71.89mAs
- Enh. Scan Std Dose: 110kV 38.14mAs - Enh. Scan Boosted Dose: 110kV 97.61mAs
38
Loading factors: automatically set by software
Eco Scan Std Dose: 110kV 23.14mAs - Reg. Scan Std Dose: 110kV 36.00mAs - Reg. Scan Boosted Dose: 110kV 64.21mAs
- Enh. Scan Std Dose: 110kV 46.28mAs - Enh. Scan Boosted Dose: 110kV 85.61mAs
39
Loading factors: automatically set by software
Eco Scan Std Dose: 110kV 25.30mAs - Reg. Scan Std Dose: 110kV 39.55mAs - Reg. Scan Boosted Dose: 110kV 71.25mAs
- Enh. Scan Std Dose: 110kV 50.60mAs - Enh. Scan Boosted Dose: 110kV 93.27mAs
40
Loading factors: automatically set by software
Eco Scan Std Dose: 110kV 72.60mAs - Reg. Scan Std Dose: 110kV 108.90mAs - Enh. Scan Std Dose: 110kV 147.01mAs
41
Loading factors: automatically set by software
Eco Scan Std Dose: 110kV 72.60mAs - Reg. Scan Std Dose: 110kV 108.90mAs - Enh. Scan Std Dose: 110kV 147.01mAs
42
Loading factors: automatically set by software
Eco Scan Std Dose: 110kV 72.60mAs - Reg. Scan Std Dose: 110kV 108.90mAs - Enh. Scan Std Dose: 110kV 147.01mAs
43
Loading factors: automatically set by software
Eco Scan Std Dose: 110kV 72.60mAs - Reg. Scan Std Dose: 110kV 89.84mAs - Enh. Scan Std Dose: 110kV 147.01mAs
Veterinary Application
10-19
Summary of Contents for NewTom 5G Vet
Page 1: ...0051 97050812 Rev 1 28 10 2016 ATTACHMENT to the User Manual Veterinary Application Attachment...
Page 38: ...10 APPENDIX A Technical references 10 8 Veterinary Application...
Page 39: ...10 APPENDIX A Technical references Veterinary Application 10 9...
Page 40: ...10 APPENDIX A Technical references 10 10 Veterinary Application...
Page 41: ...10 APPENDIX A Technical references Veterinary Application 10 11...