10 APPENDIX A- Technical references
CTDI
W
table declared for operative modalities of the device
Values expressed in mGy. Tolerance is
30%. The values correspond to CTDIw CTDIvol for this application
Values calculated according to IEC 60601-2-44: 2009 Par 201.3.211 and 201.3.212 c)
FOV
Eco Scan
Standard Dose
Regular Scan
Standard Dose
Regular Scan
Boosted Dose
Enhanced Scan
Standard Dose
Enhanced Scan
Boosted Dose
[18x16]
3,0
4,5
9,1
6,1
12,0
[18x22e]
4,2
6,4
10,0
-
-
[15x12]
3,1
4,5
9,2
6,2
12,2
[12x8]
3,4
4,9
8,3
6,8
11,0
[8x8]
2,9
4,4
7,1
5,9
9,5
[15x5] HiRes
13,4
19,6
-
26,5
-
[12x8] HiRes
10,0
15,1
-
20,4
-
[8x8] HiRes
8,3
12,5
-
16,9
-
[6x6] HiRes
8,5
10,4
-
17,2
-
Date: 2013-07-11
22
Measured by using “head phantom” according to IEC 60601-2-44: 2009 Par. 203.108
23
For [15x22e] FOV please refer to the following table
24
Loading factors: automatically set by software
Eco Scan Std Dose: 110kV 17.76mAs - Reg. Scan Std Dose: 110kV 26.35mAs - Reg. Scan Boosted Dose: 110kV 67.20mAs
- Enh. Scan Std Dose: 110kV 35.52mAs - Enh. Scan Boosted Dose: 110kV 87.64mAs
25
Loading factors: automatically set by software
Eco Scan Std Dose: 110kV 32,85mAs - Reg. Scan Std Dose: 110kV 48,84mAs - Reg. Scan Boosted Dose: 110kV 97,63mAs
26
Loading factors: automatically set by software
Eco Scan Std Dose: 110kV 19.07mAs - Reg. Scan Std Dose: 110kV 28.37mAs - Reg. Scan Boosted Dose: 110kV 71.89mAs
- Enh. Scan Std Dose: 110kV 38.14mAs - Enh. Scan Boosted Dose: 110kV 97.61mAs
27
Loading factors: automatically set by software
Eco Scan Std Dose: 110kV 23.14mAs - Reg. Scan Std Dose: 110kV 36.00mAs - Reg. Scan Boosted Dose: 110kV 64.21mAs
- Enh. Scan Std Dose: 110kV 46.28mAs - Enh. Scan Boosted Dose: 110kV 85.61mAs
28
Loading factors: automatically set by software
Eco Scan Std Dose: 110kV 25.30mAs - Reg. Scan Std Dose: 110kV 39.55mAs - Reg. Scan Boosted Dose: 110kV 71.25mAs
- Enh. Scan Std Dose: 110kV 50.60mAs - Enh. Scan Boosted Dose: 110kV 93.27mAs
29
Loading factors: automatically set by software
Eco Scan Std Dose: 110kV 72.60mAs - Reg. Scan Std Dose: 110kV 108.90mAs - Enh. Scan Std Dose: 110kV 147.01mAs
30
Loading factors: automatically set by software
Eco Scan Std Dose: 110kV 72.60mAs - Reg. Scan Std Dose: 110kV 108.90mAs - Enh. Scan Std Dose: 110kV 147.01mAs
31
Loading factors: automatically set by software
Eco Scan Std Dose: 110kV 72.60mAs - Reg. Scan Std Dose: 110kV 108.90mAs - Enh. Scan Std Dose: 110kV 147.01mAs
32
Loading factors: automatically set by software
Eco Scan Std Dose: 110kV 72.60mAs - Reg. Scan Std Dose: 110kV 89.84mAs - Enh. Scan Std Dose: 110kV 147.01mAs
Veterinary Application
10-17
Summary of Contents for NewTom 5G Vet
Page 1: ...0051 97050812 Rev 1 28 10 2016 ATTACHMENT to the User Manual Veterinary Application Attachment...
Page 38: ...10 APPENDIX A Technical references 10 8 Veterinary Application...
Page 39: ...10 APPENDIX A Technical references Veterinary Application 10 9...
Page 40: ...10 APPENDIX A Technical references 10 10 Veterinary Application...
Page 41: ...10 APPENDIX A Technical references Veterinary Application 10 11...