![Cedrat Technologies CCBu40 Product And Warranty Information Download Page 9](http://html1.mh-extra.com/html/cedrat-technologies/ccbu40/ccbu40_product-and-warranty-information_2578811009.webp)
9 / 32
CCBU40 COMPACT CONTROLLER BOARD - PRODUCT AND WARRANTY INFORMATION
Version 1.21
can then be read as a voltage on the “T°C” analog output of the supervisor connector. The CCBu40 integrates
a conditioner that provides a constant 1.613mA current to the PT1000 probe. Based on the voltage
measurement on the “T°C” output, the temperature in °C is computed as follows:
Temperature [°C] = (“T°C” output [V] – 1.613) x 161
Notes:
•
When monitoring the sensor signals on the SX and SY outputs, it is recommended to use an
low pass filter to remove switching noise that can appear on those lines; This is also
recommended for temperature reading (if used) on the T°C.