MANUFACTURER’S DECLARATION 

ELECTROMAGNETIC IMMUNITY 

TEST SPECIFICATIONS FOR ENCLOSURE PORT IMMUNITY TO RF WIRELESS COMMUNICATIONS EQUIPMENT

The THD2FE is intended for use in the electromagnetic environment (for home healthcare) specified below. The customer or the user 

of the THD2FE should assure that it is used in such an environment.

Test frequency 

(MHz)

Band 

a)

 

(MHz)

Service 

a)

Modulation 

b)

Maximum power

 

(W)

Distance 

(m)

Immunity Test Level 

(V/m)

Compliance Level 

(V/m)

(for home healthcare)

385

380–390

TETRA 400

Pulse modulation 

b)

 18 Hz

1,8

0,3

27

27

450

430–470

GMRS 460,  

FRS 460

FM 

c)

 

±5 kHz deviation 

1 kHz sine

2

0,3

28

28

710

704–787

LTE Band 13,17

Pulse modulation 

b)

 

217 Hz

0,2

0,3

9

9

745

780

810

800–960

GSM 800/900, TETRA 
800, iDEN 820, CDMA 

850, LTE Band 5

Pulse modulation 

b)

  

18 Hz

2

0,3

28

28

870
930

1720

1700–1990

GSM 1800; CDMA 1900; 

GSM 1900; DECT; LTE 

Band 1, 3, 4, 25; UMTS

Pulse modulation 

b)

 

217 Hz

2

0,3

28

28

1845
1970

2450

2400–2570

Bluetooth, WLAN, 802.11 

b/g/n, RFID 2450, LTE 

Band 7

Pulse modulation 

b)

 

217 Hz

2

0,3

28

28

5240

5100–5800

WLAN 802.11 

a/n

Pulse modulation 

b)

 

217 Hz

0,2

0,3

9

9

5500

5785

NOTE 

If necessary to achieve the IMMUNITY TEST LEVEL, the distance between the transmitting antenna and the ME EQUIPMENT or ME SYSTEM may be reduced to 1 m. The 1 m test distance is permitted by 
IEC 61000-4-3.

a) For some services, only the uplink frequencies are included. b) The carrier shall be modulated using a 50 % duty cycle square wave signal. c) As an alternative to FM modulation, 50 % pulse modulation at 18 Hz 
may be used because while it does not represent actual modulation, it would be worst case.