SYMBOL DESCRIPTIONS

Caution

BF type applied part

IP22

Classification for water ingress and 

particulate matter

Please read the instructions for use

Power, standby

Paper recycling

MANUFACTURER’S DECLARATION 

ELECTROMAGNETIC IMMUNITY

The THD2FE is intended for use in the electromagnetic 

environment (for home healthcare) specified below.  

The customer or the user of the THD2FE should assure  

that it is used in such an environment.

Immunity Test

IEC 60601  

Test Level

Compliance Level

Electromagnetic Environment – 

Guidance

 

(for home healthcare environment)

Radiated RF
IEC 61000-4-3

10 V/m
80 MHz – 2,7 GHz
80 % AM at 1 kHz

10 V/m
80 MHz – 2,7 GHz
80 % AM at 1 kHz

Recommended separation 

distance:

d = 1,2  √

P

 

d = 1,2  √

P

 

80MHz to 800 MHz

d = 2,3  √

P

 

800MHz to 2,7 GHz

Where 

P

 is the maximum output 

power rating of the transmitter 
in watts (W) according to the 
transmitter manufacturer and 

d

 

is the recommended separation 
distance in meters (m).
Field strengths from fixed RF 
transmitters, as determined by 
an electromagnetic site survey,

a

 

should be less than the compliance 
level in each frequency range.

b

 

Interference may occur in the 
vicinity of equipment marked with 
the following symbol: 

 

 

NOTE 1  At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2  These guidelines may not apply in all situations. Electromagnetic propagation is affected 

by absorption and reflection from structures, objects and people.

Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) 
telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast 
cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due 
to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured 
field strength in the location in which the THD2FE is used exceeds the applicable RF compliance 
level above, the THD2FE should be observed to verify normal operation. If abnormal performance is 
observed, additional measures may be necessary, such as re-orienting or relocating the THD2FE.

MANUFACTURER’S DECLARATION 

ELECTROMAGNETIC EMISSIONS

The THD2FE is intended for use in the electromagnetic 

environment (for home healthcare) specified below.  

The customer or the user of the THD2FE should assure  

that it is used in such an environment.

Emissions Test

Compliance

Electromagnetic Environment – Guidance  

(for home healthcare environment)

RF emissions 
CISPR 11

Group 1

The THD2FE uses RF energy only for its internal function. 
Therefore, its RF emissions are very low and are not likely 
to cause any interference in nearby electronic equipment.

RF emissions 
CISPR 11

Class B

The THD2FE is suitable for use in all establishments, 
including domestic establishments and those directly 
connected to the public low-voltage power supply 
network that supplies buildings used for domestic 
purposes.

MANUFACTURER’S DECLARATION 

ELECTROMAGNETIC IMMUNITY

The THD2FE is intended for use in the electromagnetic 

environment (for home healthcare) specified below.  

The customer or the user of the THD2FE should assure  

that it is used in such an environment.

Immunity Test

IEC 60601 Test Level

Compliance Level

Electromagnetic 

Environment – 

Guidance

 

(for home healthcare 

environment)

Electrostatic  
discharge (ESD)
IEC 61000-4-2

Contact:±8 kV
Air±2 kV,±4 kV,±8
kV,±15 kV

Contact:±8 kV
Air±2 kV,±4 kV,±8
kV,±15 Kv

Floors should be wood, 
concrete or ceramic 
tile. If floors are 
covered with synthetic 
material, the relative 
humidity should be at 
least 30%.

Power frequency 
(50, 60 Hz)  
magnetic field  
IEC 61000-4-8

30 A/m
50 Hz or 60 Hz

30 A/m
50 Hz and 60 Hz

The THD2FE power 
frequency magnetic 
fields should be at 
levels characteristic 
of a typical 
location in a typical 
home healthcare 
environment.

RECOMMENDED SEPARATION DISTANCES  

BETWEEN PORTABLE AND MOBILE RF 

COMMUNICATIONS EQUIPMENT AND THE THD2FE

The THD2FE is intended for use in an electromagnetic 

environment (for home healthcare) in which radiated RF 

disturbances are controlled. The customer or the user of the 

THD2FE can help prevent electromagnetic interference by 

maintaining a minimum distance between portable and mobile 

RF communications equipment (transmitters) and the THD2FE 

as recommended below, according to the maximum output 

power of the communications equipment.

Rated maximum 

output power of 

transmitter

 

W

Separation distance according to frequency of transmitter 

m

150 kHz to 80 MHz

 

d

 = 1,2 √

P

80 MHz to 800 MHz

 

d

 = 1,2 √

P

800 MHz to 2,7 GHz 

d

 = 2,3 √

P

0,01

N/A

0,12

0,23

0,1

N/A

0,38

0,73

1

N/A

1,2

2,3

10

N/A

3,8

7,3

100

N/A

12

23

For transmitters rated at a maximum output power not listed above, the recommended separation 
distance 

d

 in meters (m) can be estimated using the equation applicable to the frequency of the 

transmitter, where 

P

 is the maximum output power rating of the transmitter in watts (W) according 

to the transmitter manufacturer.
NOTE 1  At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2  These guidelines may not apply in all situations. Electromagnetic propagation is affected 

by absorption and reflection from structures, objects and people.