SYMBOL DESCRIPTIONS
Caution
BF type applied part
IP22
Classification for water ingress and
particulate matter
Please read the instructions for use
Power, standby
Paper recycling
MANUFACTURER’S DECLARATION
ELECTROMAGNETIC IMMUNITY
The THD2FE is intended for use in the electromagnetic
environment (for home healthcare) specified below.
The customer or the user of the THD2FE should assure
that it is used in such an environment.
Immunity Test
IEC 60601
Test Level
Compliance Level
Electromagnetic Environment –
Guidance
(for home healthcare environment)
Radiated RF
IEC 61000-4-3
10 V/m
80 MHz – 2,7 GHz
80 % AM at 1 kHz
10 V/m
80 MHz – 2,7 GHz
80 % AM at 1 kHz
Recommended separation
distance:
d = 1,2 √
P
d = 1,2 √
P
80MHz to 800 MHz
d = 2,3 √
P
800MHz to 2,7 GHz
Where
P
is the maximum output
power rating of the transmitter
in watts (W) according to the
transmitter manufacturer and
d
is the recommended separation
distance in meters (m).
Field strengths from fixed RF
transmitters, as determined by
an electromagnetic site survey,
a
should be less than the compliance
level in each frequency range.
b
Interference may occur in the
vicinity of equipment marked with
the following symbol:
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected
by absorption and reflection from structures, objects and people.
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless)
telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast
cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due
to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured
field strength in the location in which the THD2FE is used exceeds the applicable RF compliance
level above, the THD2FE should be observed to verify normal operation. If abnormal performance is
observed, additional measures may be necessary, such as re-orienting or relocating the THD2FE.
MANUFACTURER’S DECLARATION
ELECTROMAGNETIC EMISSIONS
The THD2FE is intended for use in the electromagnetic
environment (for home healthcare) specified below.
The customer or the user of the THD2FE should assure
that it is used in such an environment.
Emissions Test
Compliance
Electromagnetic Environment – Guidance
(for home healthcare environment)
RF emissions
CISPR 11
Group 1
The THD2FE uses RF energy only for its internal function.
Therefore, its RF emissions are very low and are not likely
to cause any interference in nearby electronic equipment.
RF emissions
CISPR 11
Class B
The THD2FE is suitable for use in all establishments,
including domestic establishments and those directly
connected to the public low-voltage power supply
network that supplies buildings used for domestic
purposes.
MANUFACTURER’S DECLARATION
ELECTROMAGNETIC IMMUNITY
The THD2FE is intended for use in the electromagnetic
environment (for home healthcare) specified below.
The customer or the user of the THD2FE should assure
that it is used in such an environment.
Immunity Test
IEC 60601 Test Level
Compliance Level
Electromagnetic
Environment –
Guidance
(for home healthcare
environment)
Electrostatic
discharge (ESD)
IEC 61000-4-2
Contact:±8 kV
Air±2 kV,±4 kV,±8
kV,±15 kV
Contact:±8 kV
Air±2 kV,±4 kV,±8
kV,±15 Kv
Floors should be wood,
concrete or ceramic
tile. If floors are
covered with synthetic
material, the relative
humidity should be at
least 30%.
Power frequency
(50, 60 Hz)
magnetic field
IEC 61000-4-8
30 A/m
50 Hz or 60 Hz
30 A/m
50 Hz and 60 Hz
The THD2FE power
frequency magnetic
fields should be at
levels characteristic
of a typical
location in a typical
home healthcare
environment.
RECOMMENDED SEPARATION DISTANCES
BETWEEN PORTABLE AND MOBILE RF
COMMUNICATIONS EQUIPMENT AND THE THD2FE
The THD2FE is intended for use in an electromagnetic
environment (for home healthcare) in which radiated RF
disturbances are controlled. The customer or the user of the
THD2FE can help prevent electromagnetic interference by
maintaining a minimum distance between portable and mobile
RF communications equipment (transmitters) and the THD2FE
as recommended below, according to the maximum output
power of the communications equipment.
Rated maximum
output power of
transmitter
W
Separation distance according to frequency of transmitter
m
150 kHz to 80 MHz
d
= 1,2 √
P
80 MHz to 800 MHz
d
= 1,2 √
P
800 MHz to 2,7 GHz
d
= 2,3 √
P
0,01
N/A
0,12
0,23
0,1
N/A
0,38
0,73
1
N/A
1,2
2,3
10
N/A
3,8
7,3
100
N/A
12
23
For transmitters rated at a maximum output power not listed above, the recommended separation
distance
d
in meters (m) can be estimated using the equation applicable to the frequency of the
transmitter, where
P
is the maximum output power rating of the transmitter in watts (W) according
to the transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected
by absorption and reflection from structures, objects and people.