3

c. With 

UNIT 

appearing on the display, press the 

UNIT button. 

AOFF

 appears on the display.

d. Press the SET button to begin selection of  

the auto off time (in seconds). 

120

 appears  

on the display.

e. Press the UNIT button to cycle through the 

selections: 

120

180

240

300

 or 

OFF

.

f.  When the desired time is displayed, press the 

SET button to confirm the selection. 

AOFF

 

appears on the display.

g. Press the UNIT button twice. 

END

 appears on 

the display. 

h. Press the SET button to save the settings and 

return to weighing mode.

3. Gravity Compensation  

The gravity compensation for your scale has 
been pre-configured at the factory and should 
not require changing in most locations. In the 
event that the factory settings do not meet the 
requirements of your location, contact CDN for 
the value for your location. With 

GrAu

 enabled, 

the scale will use the default value of 9.797. If 
the 

GrAu

 is disabled, the scale will use the value 

obtained from CDN and entered by the operator 
for the gravity compensation.

a. Enabling/Disabling Gravity Compensation

1) Press the POWER button ( ) to turn on.
2) Press and hold the SET button for  

3 seconds. When 

SEt 

appears on the 

display, release the SET button to enter  
the configuration mode. 

UNIT

 appears on 

the display.

3) With 

UNIT 

appearing on the display, press 

the UNIT button twice. 

GrAu

 appears on  

the display.

4) With 

GrAu 

appearing on the display, press 

the SET button again to enable the gravity 
compensation. 

ON

 appears on the display.

5) Press the UNIT button to toggle between 

ON 

(enabled) and 

OFF

 (disabled).

6) Press the UNIT button again to disable  

the gravity compensation. 

OFF

 appears  

on the display.

7) Press the SET button to confirm the 

selection. 

GrAu

 appears on the display.

8) Press the UNIT button. 

END

 appears on  

the display. 

9) Press the SET button to save the settings 

and return to weighing mode.

b. Gravity Compensation Value Editing 

Contact CDN for the value to enter for  
your location. 

Note:

 Gravity compensation must be disabled for this 

value to take effect.

1) Press and hold the UNIT button and press 

the POWER button ( ) to turn on the scale 
and enter the value editing mode. 

9.797

 (or 

the current value) appears on the display.

2) Press the TARE button to choose a number 

from 0 to 9 to enter the desired gravity value.

3) Press the UNIT button to move to the next 

digit to the right.

4) Repeat steps 2 and 3 for the  

remaining digits.

5) Press the SET button to confirm the entered 

gravity value and return the scale to the 
normal operating mode.

Note:

 Clean the scale platform before each use.

Note: 

Replacement AC adapter, removable platform & 5 kg 

calibration weight available.

USAGE GUIDELINES

• Scales are precision instruments and must be 

handled with extreme care.

• Battery technology works best at room temperature.
• Place the scale gently on a hard and flat surface and 

make sure  
it is steady before use.

• The maximum capacity of the scale is 11 lb/5 kg, 

which is the capacity before using the tare function.

• Do not place overweight items on the scale.
• Store the scale in a position that keeps it free from 

any load bearing weight. Storing the scale with 
pressure on the load cells can distort the weight 
measurements over time.

• Hold the scale from the bottom when moving.

CARE OF YOUR PRODUCT

• 

IP67

 — protected from water up to 3 feet for  

30 minutes, completely protected against dust.

• Wipe clean with a damp cloth.
• Store the scale at room temperature between  

32 to 104°F/0 to 40°C.

Summary of Contents for SD1110

Page 1: ...in and does not need to be replaced 1 Insert the adapter plug into the jack on the back of the scale Allow the scale to charge for 5 hours The scale will then operate continuously for approximately 14 hours Operating Instructions Place the scale gently on a hard and flat surface and make sure it is steady before use to obtain accurate measurements A On Off 1 Press the POWER button to turn on The d...

Page 2: ... before automatically turning off The calibration is complete H Battery Status Low Battery Battery status is low and needs to be charged See Charging the Battery on page 1 I Error Messages oL indicates an overload when the item being measured exceeds the capacity of the scale Immediately remove the overweight item to avoid causing permanent damage to your scale J Programmable Modes The scale has b...

Page 3: ...F appears on the display 7 Press the SET button to confirm the selection GrAu appears on the display 8 Press the UNIT button END appears on the display 9 Press the SET button to save the settings and return to weighing mode b Gravity Compensation Value Editing Contact CDN for the value to enter for your location Note Gravity compensation must be disabled for this value to take effect 1 Press and h...

Page 4: ...ic discharge or malfunctioning occurs please re install the battery to reset this unit 10 2016 Component Design Northwest Inc Made in China CD9999319en 10 16 L DESIGN 614 525 1472 The information in this document has been reviewed and is believed to be accurate However neither the manufacturer nor its affiliates assume any responsibility for inaccuracies errors or omissions that may be contained h...

Reviews: