4.5 Delete Memory
Selecting “Delete Memory” from the Utility Menu, all the measure data and flow records are
permanently erased from the memory.
4.6 Measure Test and Calibrations
•
From the Utility Menu, selecting Measures
Calibrations a clock glass and a message asking you
to wait will appear on the display.
The temperature values displayed are the temperature sensor readings.
•
Tc.v. = Dry Gas Meter Temperature
•
Pressing the
1
key, it’s possible to enter the
temperature value measured by the reference
instrument.
•
Pressing
ENTER
you’ll go back to the previous menu.
The calibration of those parameters requires specific equipment (temperature
calibrators and tools).
Attention: if the calibrator reported value differs a lot (more than 20%) from the
instrument value, do not proceed with the calibration and contact the CDL-Tecora
Service or your local dealer.
UTILITY MENU
Test EV
OFF
Set datary
Baud 9600
Download Memory
Delete Memory
Measures Calibrations
Load Default Calibrations
ESC
Sensors Measure Test
Tc.v. °C
23.055
--1--
Calibrate
ESC