Check the relays and all components subject to wear. For every replacement use original
spare parts only.
Check the conditions of the pipe carrying the aspirated liquid; especially the portion
after the draining pump; in case of cracks, however small, pipes must be replaced.
See that the staff in charge of cleaning and disinfection carries out maintenance duly
and using the recommended products: it is the responsibility of the technician in
charge of general maintenance to instruct the surgery staff as to the requested daily
maintenance operations.
Appliances or their functioning must never be modifi ed.
Maintenance entrusted to the surgery staff
• Electrical shock risk: also 230 V
can be lethal
• Biological danger, risk of infections from epidemic diseases
Follow the recommendations in the paragraph “General warnings and biological risk” and
the recommendations for technical checking.
According to the instructions, at the end of every surgical and particularly long surgical
operation rinse the system thoroughly aspirating clear water. At the end of every working
day aspirate a solution of Puli-Jet plus and hot water (+ 50°C) using Pulse Cleaner to
cleanse and disinfect the equipment and replace dirty fi lters. In the morning, before the
fi rst treatment, check that the Sanitizing Antifoam Tablets have been put in the debris
fi lter. Using products not recommended by the manufacturer can jeopardize the
operation of the equipment and cause damages.
Any repair and intervention on the machine must be carried out by a skilled technician
authorized by the manufacturer.
Important notices
• Appliances are guaranteed for one year from the date of sale, provided that the warranty
slip addressed to the manufacturer is returned to the manufacturer with date of sale,
retailer's and customer's name.
• Warranty and manufacturer’s liability cease in case appliances are treated with products
which are unsuitable or different from those recommended by the manufacturer and also
in case appliances are improperly used or tampered with operations of any kind carried
out by people who are not authorized by the manufacturer.
• The manufacturer, concessionaires, agents and authorized technicians are at customers'
disposal for advice and assistance and to supply literature, spare parts and anything
useful.
• The manufacturer reserves the right to modify the products for improvements, for techni-
cal, normative and functional reasons or for problems due to the availability of products
or semi-fi nished products, without prior notice.
• Our updated manuals are available on the web site www.cattani.it.
We recommend they are consulted especially for updates about safety.
Transport of second hand appliances
• Before packing, cleanse and disinfect with PULI-JET Plus (refer to the paragraph
"Maintenance entrusted to the surgery staff”)
• Place the machine into a polyethylene bag, seal and pack it in 3-layer corrugated
board.