Technical features
Rated voltage
24 V
AC
Frequency 50/60
Hz
Power 15
W
Fuse
T 2,5 A
Protection Class
Alternating current
IEC 417-5032
Earthing (ground)
IEC 417-5019
Installation
(Draw. 1)
The appliance is to be installed inside the dental unit or inside
a floor box next to the dental unit (020996 and 020998).
In the first case the Mini-Separator instructions are
included in the dental unit manual.
For the installation of the box fitting
the Mini-Separator refer to the
manual "Box for Maxi-Canister/Mini-Separator".
Functioning
(Draw. 1)
The aspirated fluid must be filtered through a 0,8 mm
filter and conveyed to the Mini-Separator (Draw. 1, det. A)
through a slightly sloping pipe to facilitate its function-
ing. Air comes out from the outlet (B), whereas the
liquid collects on the bottom of the tank.
When the liquid level reaches the two
longest probes (C) the draining pump
(D) conveys the liquid to waste (E).