28 |
P a g e
cascadesciences.com
|
Tel. 503 847-9047
INSTALLATION
C
ONNECT TO THE
V
ACUUM
S
UPPLY
•
The Vacuum Port – 3/8 inch (9.52mm) OD
o
Connect a vacuum pump or building vacuum supply system to this port. Chamber
atmosphere is evacuated through this port.
o
This port is opened and closed by the Vacuum Valve control on the front control
panel.
•
The Chamber Vent Port (Backfilling) – 1/4 inch (6.35mm) OD
o
This intake port allows external atmosphere to backfill the oven chamber when the
Vent Valve control on the front control panel is turned to open.
o
Optional
: A clean or inert gas supply source may be connected to this port. Please
page 29.
•
KF-25 Fitting
o
This large fitting is normally used to introduce sensor probes into the oven chamber
using a vacuum feed through.
o
Optional
: A vacuum supply can be connected to the KF-25 fitting for increased
efficiency in vacuuming down the chamber. However, the Vacuum Valve control on
the front control panel will not affect the level of vacuum and must be set to closed
to prevent atmosphere from entering the chamber through the 3/8” Vacuum Port.
Vacuum Port
3/8 inch
(9.52mm)
Vacuum Connections
KF-25
Vacuum
Flange
Use of clamps to secure tubing to the Vacuum Port and Chamber Vent is recommended.
Vacuum Line Connected to the
Vacuum Port
Summary of Contents for CVO-2
Page 1: ...Installation Operation Manual CVO 2 Benchtop Vacuum Ovens...
Page 6: ...6 P a g e cascadesciences com Tel 503 847 9047 TABLE OF CONTENTS...
Page 8: ...8 P a g e cascadesciences com Tel 503 847 9047 CERTIFICATIONS...
Page 12: ...12 P a g e cascadesciences com Tel 503 847 9047 SPECIFICATIONS...
Page 20: ...20 P a g e cascadesciences com Tel 503 847 9047 RECEIVING...
Page 30: ...30 P a g e cascadesciences com Tel 503 847 9047 INSTALLATION...
Page 32: ...32 P a g e cascadesciences com Tel 503 847 9047 SYMBOLS...
Page 50: ...50 P a g e cascadesciences com Tel 503 847 9047 MAINTENANCE...
Page 52: ......