CONTROL
Vacuum Gauge
This gauge is set at the factory to show the chamber vacuum level relative to sea level
atmospheric pressure in inches of mercury (inHg). The display range is 0 to -29.9inHg. Zero
is the room atmosphere pressure at sea level and -29.9inHg a near-perfect vacuum. See
page 44 for how to display other units of measurement or zero the gauge to your local
altitude.
Vacuum Valve Control
This valve adjusts the level of vacuum draw applied to the oven chamber through
the vacuum port on the back of the oven.
•
When
open
, this valve allows the connected vacuum supply to evacuate the oven
chamber.
•
In the
closed
position, the valve cuts off the vacuum draw.
Vent Valve Control – Backfill Inlet
This valve controls the chamber inlet Vent port on the back of the oven.
•
In the
open
position, the oven chamber is open to external atmosphere through
the vent intake port on the back of the oven.
•
Optional: An inert or clean backfilling gas supply connected to the Vent Port will
flow gas from the pressurized supply to the oven chamber when the Vent Valve is
open.
•
When the valve control is in the
closed
position, the chamber is cut off from
external atmosphere and any gas supply.
o
The vent must be closed prior to vacuuming down the chamber.
Failure to do so may result in damage to your vacuum pump.
Closed
Open
Summary of Contents for CVO-2
Page 1: ...Installation Operation Manual CVO 2 Benchtop Vacuum Ovens...
Page 6: ...6 P a g e cascadesciences com Tel 503 847 9047 TABLE OF CONTENTS...
Page 8: ...8 P a g e cascadesciences com Tel 503 847 9047 CERTIFICATIONS...
Page 12: ...12 P a g e cascadesciences com Tel 503 847 9047 SPECIFICATIONS...
Page 20: ...20 P a g e cascadesciences com Tel 503 847 9047 RECEIVING...
Page 30: ...30 P a g e cascadesciences com Tel 503 847 9047 INSTALLATION...
Page 32: ...32 P a g e cascadesciences com Tel 503 847 9047 SYMBOLS...
Page 50: ...50 P a g e cascadesciences com Tel 503 847 9047 MAINTENANCE...
Page 52: ......