8
2. Calibration Mode
To go to calibration mode, turn on the power while pressing the calibration switch.
The display shows “CAL” three times, and then “C-0”.
Mode Selection : [TARE] or [UNIT] key
Enter : [ZERO] key
MODE
DESCRIPTION
C – 0
Normal Mode
C – 1
Span calibrated A/D value
C – 3
Span Calibration
C – 4
Capacity Display and Option Setting
C – 5
Averaging A/D Value
C – 6
Real A/D Value
C – 7
% Calibration (10%~100%)
C – 9
Gravity Constant
Calibration Mode
MODE
DESCRIPTION
C-4-1
Save Setting
Calibration Unit, Tare setting
C-4-2
Capacity Setting
Capacity setting
C-4-3
Comma Setting
Comma setting
C-4-4
Capacity Setting
Save setting
C – 4 Setting