2.2. Span Calibration(C – 3)
(1) To go to calibration mode, turn on the power while pressing the calibration switch. The
display shows “CAL” three times, and then “C-0”.
(2) Press the [TARE] key until the display shows “C – 3”.
(3) Press the [ZERO] key, the display shows “UnLOAd”.
(4) Remove items from the platter and press the [ZERO] key. The display shows “StAbLE”
and then “LOAd”. (If you want to finish this mode, press the TARE key.)
(5) Place full capacity on the platter and press the [ZERO] key. The display shows “C-3End”
and “C-3”.
(6) Remove full capacity from the platter. You can change the mode by pressing TARE or
lb/kg(*) key.