51
Err 11 (at Span calibration)
■
Cause
The weight for span calibration is greater than 100% of the maximum capacity.
■
Management
The weight for span calibration should be within 10~100% of the maximum
capacity in ‘BAL’.
Err 12 (at Span calibration)
■
Cause
The weight for span calibration is lower than 10% of the maximum capacity.
■
Management
The weight for span calibration should be within 10~100% of the maximum
capacity in ‘BAL’.
Err 13 (Zero Range when the Power is ON)
■
Cause
The initial zero range exceeds above 10% of maximum capacity
■
Management
Unload the scale
Check the A/D value (at ‘TEST 3’).
If the load cell is damaged, replace the load cell
Err 14 (at Span calibration)
■
Cause
The weight for span calibration is 0kg.
■
Management
The weight for span calibration should be within 10~100% of the maximum
capacity in ‘BAL’.
BAT or LobAt
Summary of Contents for NC-1
Page 1: ...2 ...
Page 20: ...21 11 Select hex file 12 Download is done 13 If download is done you must turn off SW5 FIG 2 ...
Page 24: ...4 Cal schematic SCHEMATIC DIAGRAM 25 ...
Page 25: ...26 8 PCB BOARD DIAGRAM 1 Main PCB board LED ...
Page 26: ...27 2 Display PCB board ...
Page 27: ...28 3 Charger PCB board ...
Page 28: ...29 9 DEVICE OF SPECIFICATION 1 MPU ADUC845 ...
Page 29: ...30 ...
Page 30: ...31 ...
Page 31: ...32 ...
Page 32: ...33 2 OPAMP OP2177 OP2277 ...
Page 33: ...34 ...
Page 34: ...35 ...
Page 35: ...36 3 REGULATOR LM2575 ...
Page 36: ...37 ...
Page 37: ...38 ...
Page 38: ...39 ...
Page 39: ...40 ...
Page 40: ...41 ...
Page 41: ...42 ...
Page 42: ...43 4 RESET TC1274 ...
Page 43: ...44 ...
Page 44: ...45 ...
Page 45: ...46 ...
Page 46: ...47 4 RESET TC1274 ...
Page 47: ...48 ...
Page 48: ...49 ...
Page 82: ...83 12 EXPLODE VIEW ...
Page 83: ...84 ...