71
11. Error codes
During the initial power-on testing it is possible the scale may show error message.
The meaning of the error messages is described below.
ERROR CODE
POSSIBLE CAUSE
HANDLING
E1
The scale hasn’t be calibrated
Before or calibration data lost.
Calibrate the scale.
E2
EPROM data lost.
Recalibrate the scale.
Remote platform is not well
connected with the scale when
powers on.
Connect the remote platform
properly and switch on again.
E3
.
1. Local platform is not placed well.
2. There are something heavy
touch the pan.
1. Place the pan well and
switch on again.
2. Remove the weight and
switch on again.
E4
Address code of Unit Weight is out
of “1 ~ 200”.
Correct the operation.
E5
In alarm setting, the LO value is set
higher than HI value.
Correct the operation.
OL
Overload
Take off the weight
immediately.
Low battery
Charge the battery.
If the error message is still shown after above ways, please recalibrate. If the problem still
Can not be solved then contact your dealer for further support.