1. Precautions Before Using The Scale
Environment
The scale should always be used in an environment, which is free from excessive
air currents, corrosives, vibration, and temperature or humidity extremes.
These factors will affect displayed weight reading.
DO NOT
install the scale:
●
Next to open windows or doors causing drafts or rapid temperature changes.
●
Near air conditioning or heating vents.
●
Near vibrating, rotating or reciprocating equipment.
●
Near magnetic fields or equipment that generates magnetic fields
.
●
On an unstable work surface
●
In a dusty environment
●
In direct sunlight.
Leveling the Scale
The scale is equipped with a level indicator on the back side, right bottom of the
front panel and four adjustable leveling feet. Adjust the leveling feet until the
bubble appears in the center circle of the indicator.
Turn on Scale
Do not turn on scale with anything on the platform.
Press the “ON/OFF” switch located on the right side of the bottom of the scale
to turn on the scale.
The scale will start to count down from nine to zero. The scale is then ready for use.
Give a warm-up for 15~30 minutes before use.
※
Attention
※
There is a dust protection cover as standard.
Before turning on the scale, the dust protection cover should be attached on the body
with using an adhesive tape so that the cover does not touch the pan. If the cover
touches the pan, a weight value can be wrong.