- 23 -
ERROR MESSAGE & TROUBLE SHOOTING
Errors in weighing mode
Err 01
▶
Cause
The indicator can’t be initialized due to platform shaking.
▶
Management
Put the platform on a flat place without shaking and power on.
Err 02
▶
Cause
The loadcell is connected wrongly or A/D conversion part has a fault.
▶
Management
Check that the loadcell is well connected with main unit.
Err 03
▶
Cause
The calibration isn’t executed or the data of internal memory is lost
due to a electrical shock.
▶
Management
Perform the required specification again in setting mode.
Err 09
▶
Cause
The current weight exceeds zero range.
Err 10
▶
Cause
The specified TARE weight exceeds the maximum capacity.
▶
Management
Set the TARE weight lower than max. capacity.
Or set the maximum capacity higher than the desired TARE weight
in calibration menu. and perform the calibration again.
Err 13
▶
Cause
The initial zero range exceeds above 10% of maximum capacity.
Management
Check the loadcell state.
Over
▶
Cause
Too heavy object is put on platform so that it may exceed the
maximum capacity.
▶
Management
Never put the weight exceeding the maximum capacity
If the loadcell is damaged, replace the loadcell.
Summary of Contents for BW-SERIES
Page 1: ... 1 BATTERY WEIGHING SCALE BW SERIES SERVICE MANUAL Revision 2005 5 9 ...
Page 10: ... 10 SCHEMATIC DIAGRAM ...
Page 11: ... 11 P C B DIAGRAM MAIN PCB BOTTOMSIDE TOP SIDE ...
Page 12: ... 12 DISPLAY PCB BOTTOM SIDE TOP SIDE ...
Page 13: ... 13 CONNECTION DIAGRAM ...
Page 20: ... 20 BODY ASS Y LOW CAPACITY CAPACITY 30kg 15kg 6kg SPACE mm 2 4 3 8 1 35 ...
Page 21: ... 21 SET C T BOX ASS Y LOW CAPACITY ...
Page 22: ... 22 BODY ASS Y HIGH CAPACITY ...