ERROR MESSAGE DESCRIPTION & MANAGEMENT
Errors in Weighing Mode
Err 01
■
Cause
The Indicator can't be initialized due to platform shaking.
■
Management
Put the platform on a flat place without shaking and power on.
Err 02
■
Cause
The load cell is connected wrongly or A/D conversion part has a fault.
■
Management
Check that the load cell is well connected with main unit.
Err 03
■
Cause
The calibration isn't executed or the data of internal memory is lost due to
a electrical shock.
■
Management
Perform the required specification again in setting mode.
Err 09 (Zero Key)
■
Cause
The current weight exceeds zero range.
Err 10
■
Cause
The specified TARE weight exceeds the maximum capacity.
■
Management
Set the TARE weight lower than maximum capacity.
Or set the maximum capacity higher than the desired TARE weight in
calibration menu and perform the calibration again.