7340_7350-25-33 Series Operations Manual
Rev J 3-31-21
Page 56 of 92
Press the
(Temp Alarm) button.
Once the alarm screen appears, press the
(Temp High Limit) button.
Keypad screen will appear. Enter the High Temp Alarm value; press
(Enter)
when complete.
To change the Humidity Alarm, press the
(Alarms) button on the navigation
menu to go back to the Alarms screen. Press the
(Humidity Alarm) button
and repeat the same steps for humidity.
To change the alarm delay, press the alarm delay button
and enter the delay
in minutes.
Temperature
Alarm High Limit
Temperature
Alarm Low Limit
Alarm Delay