Parameter
Voltage
Current
Corresponding Value
Temperature
0 – 5 V
4-20 mA
-50 – 100 °C (adjustable*)
Humidity
0 – 5 V
4-20 mA
0 – 100 %rh
CO
2
0 – 5 V
4-20 mA
0 – 20 %CO
2
*Default range is -50C to +100C. Temperature scale low range is adjustable from -50C
to 0C. Temperature scale high range is adjustable from 1C to 100C.
Connect shielded wires to the appropriate signal terminals: I(+) for current (mA)
or
V(+)
for voltage (DC). For both current and voltage outputs, COM(-) is common terminal.
Installing Drain Water Pump (PUMP301)
In applications where a floor drain is not available and a CARON water recycling
system is not being used, a drain pump can be purchased to pump any excess
condensate from the chamber to a local sink or drain. The pump is located near the
middle of the back of the chamber. Connect the supplied tubing from the pump to the
sink / drain. The tubing may be run vertically into a ceiling but should not exceed 15
feet height. The pump is equipped with a small reservoir on the bottom of the pump
with an internal level switch that will automatically turn the pump
ON
when it is full to
drain the water out of the reservoir and into a floor or sink drain.
Pump Outlet
to Sink or Floor
Drain
Pump Inlet
from Chamber
Drain
Reservoir with
Internal Level
Switch
Summary of Contents for 7340-25
Page 1: ......
Page 80: ...7340_7350 25 33 Series Operations Manual Rev J 3 31 21 Page 80 of 92 ELECTRICAL SCHEMATICS ...
Page 81: ...7340_7350 25 33 Series Operations Manual Rev J 3 31 21 Page 81 of 92 ...
Page 82: ...7340_7350 25 33 Series Operations Manual Rev J 3 31 21 Page 82 of 92 ...
Page 83: ...7340_7350 25 33 Series Operations Manual Rev J 3 31 21 Page 83 of 92 ...
Page 84: ...7340_7350 25 33 Series Operations Manual Rev J 3 31 21 Page 84 of 92 ...
Page 85: ...7340_7350 25 33 Series Operations Manual Rev J 3 31 21 Page 85 of 92 ...
Page 86: ...7340_7350 25 33 Series Operations Manual Rev J 3 31 21 Page 86 of 92 ...
Page 87: ...7340_7350 25 33 Series Operations Manual Rev J 3 31 21 Page 87 of 92 ...
Page 88: ...7340_7350 25 33 Series Operations Manual Rev J 3 31 21 Page 88 of 92 ...