7340_7350-25-33 Series Operations Manual
Rev J 3-31-21
Page 17 of 92
Connecting the Drain Line
The chamber drain connection is located in the bottom middle of the back of the
chamber. A 3/8” tube fitting, tubing and wire ties are supplied in the unit parts kit. Insert
the tube fitting into the tubing, secure tubing to fitting with provided wire tie. Insert fitting
into drain connection. Pull on the tubing after installation to make sure it is secure. Route
the drain tubing to a local floor drain. Duplicate fitting installation on other end of tubing
if necessary
If a local floor drain is not available, a variety of accessories are available through
CARON customer service. These accessories can also be viewed at
www.caronproducts.com
.
For non-humidified chambers operated above 15°C, a drain is not needed. Install the
3/8” plug into the drain connection.
When using a pressurized water source, failure to connect
the unit to a drain could result in facility flooding.
The drain line relies on gravity to remove water from the
chamber. The drain line must remain below the chamber
to drain properly. Kinks or elevations in the drain line
above the cabinet drain will not allow the chamber to drain.
Summary of Contents for 7340-25
Page 1: ......
Page 80: ...7340_7350 25 33 Series Operations Manual Rev J 3 31 21 Page 80 of 92 ELECTRICAL SCHEMATICS ...
Page 81: ...7340_7350 25 33 Series Operations Manual Rev J 3 31 21 Page 81 of 92 ...
Page 82: ...7340_7350 25 33 Series Operations Manual Rev J 3 31 21 Page 82 of 92 ...
Page 83: ...7340_7350 25 33 Series Operations Manual Rev J 3 31 21 Page 83 of 92 ...
Page 84: ...7340_7350 25 33 Series Operations Manual Rev J 3 31 21 Page 84 of 92 ...
Page 85: ...7340_7350 25 33 Series Operations Manual Rev J 3 31 21 Page 85 of 92 ...
Page 86: ...7340_7350 25 33 Series Operations Manual Rev J 3 31 21 Page 86 of 92 ...
Page 87: ...7340_7350 25 33 Series Operations Manual Rev J 3 31 21 Page 87 of 92 ...
Page 88: ...7340_7350 25 33 Series Operations Manual Rev J 3 31 21 Page 88 of 92 ...