Caron Measurement & Controls | www.caroncontrols.net
10
S
ECTION
3.2
I
NSTALLATION
P
ROPER INSTALLATION IS INTEGRAL TO RELIABLE AND ACCURATE OPERATION
OF YOUR
T
OP
C
UT
S
AMPLER
.
T
HE FOLLOWING INFORMATION IS MEANT TO
ASSIST YOU IN SETTING UP THE SAMPLER FOR OPTIMAL OPERATION
.
F
IRST STEP IS TO FIND AN APPROPRIATE LOCATION FOR THE SAMPLE POINT
.
I
T IS
IMPORTANT TO SELECT A SAMPLE POINT WHERE THE PRODUCT WILL BE WELL
MIXED
.
W
E RECOMMEND THE USE OF A STATIC MIXER UPSTREAM OF THE
SAMPLE POINT FOR MOST APPLICATIONS
.
V
ARIOUS SIZES AND STYLES OF MIXER
ARE AVAILABLE
;
CONTACT OUR SALES DEPARTMENT FOR MIXER AND SAMPLE
QUILL OPTIONS
.
W
HEN INSTALLING THE SAMPLE QUILL
,
THE BEVELED EDGE
SHOULD FACE UPSTREAM
,
AGAINST FLOW
.
I
DEALLY THE QUILL SHOULD BE
MOUNTED OFF THE TOP OF THE PIPE
.
S
EE FIGURE
3.2.1
BELOW
.
F
IGURE
3.2.1
A
NOTHER POINT TO CONSIDER WHEN SELECTING A SAMPLE POINT LOCATION IS
THE SAMPLER ITSELF
.
F
IND A SECURE MOUNTING LOCATION FOR THE SAMPLER
AS CLOSE TO THE SAMPLE POINT AS POSSIBLE
.