CareTouch MFT-9 Manual Download Page 5

a. Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM 

and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to 
fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the MFT-9 is 
used exceeds the applicable RF compliance level above, the MFT-9 should be observed to verify normal operation. If abnormal performance is 
observed, additional measures may be necessary, such as reorienting or relocating the MFT-9. 

b. Over the frequency range 150 kHz to 80 MHz, field strengths should be less than  3  V/m. 

Guidance and manufacturer’s declaration – electromagnetic immunity 

The MFT-9 is intended for use in the electromagnetic environment specified below. The customer or the user of the MFT-9 should assure that it is 
used in such an environment. 

Immunity test

 

IEC 60601 test level

 

Compliance level

 

Electromagnetic environment 

– 

guidance

 

Electrostatic                   
discharge (ESD) 
IEC 61000-4-2 

6 kV contact 
8 kV air 

6 kV contact 
8 kV air 

Floors should be wood, concrete or ceramic tile. If floors are 
covered with synthetic material, the relative humidity should be at 
least 30 %. 

Electrical fast 
transient/burst 
 
IEC 61000-4-4 

2 kV for power supply lines 
1 kV for input/output lines 

Not applicable 

Mains power quality should be that of a typical commercial 
or hospital environment. 

Surge 
IEC 61000-4-5 

1 kV line(s) to line(s) 
2 kV line(s) to earth 

Not applicable 

Mains power quality should be that of a typical commercial 
or hospital environment. 

interruptions and 
voltage variations 
on power supply 
input lines 
 
IEC 61000-4-11 

<5 % UT 
(>95 % dip in UT) 
for 0,5 cycle 

 

40 % UT 
(60 % dip in UT) 
for 5 cycles 

 

70 % UT 
(30 % dip in UT) 
for 25 cycles 

 

<5 % UT 
(>95 % dip in UT) 
for 5 sec 

Not applicable 

Mains power quality should be that of a typical commercial 
or hospital environment. If the user of the MFT-9 requires 
continued operation during power mains interruptions, it is 
recommended that the MFT-9 be powered from an 
uninterruptible power supply or a battery. 

Power frequency 
(50/60 Hz) 
magnetic field 
 
IEC 61000-4-8 

3 A/m 

3 A/m 

Power frequency magnetic fields should be at levels 
characteristic of a typical location in a typical commercial or 
hospital environment. 

NOTE UT is the a.c. mains voltage prior to application of the test level. 

 

Recommended separation distances between 

portable and mobile RF communications equipment and the  ME EQUIPMENT or ME SYSTEM

 

The MFT-9 is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of 
the MFT-9 can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications 
equipment (transmitters) and the MFT-9 as recommended below, according to the maximum output power of the communications equipment.

 

Rated maximum output power 

of transmitter 

Separation distance according to frequency of transmitter 

m

 

150 kHz to 80 MHz 

d = 

1,2 

 

80 MHz to 800 MHz 

d = 

1,2 

 

800 MHz to 2,5 GHz 

d = 

2,3 

 

0,01

 

N/A 

0,12 

0,23 

0,1

 

N/A

 

0,38

 

0,73

 

1

 

N/A

 

1,2

 

2,3

 

10

 

N/A

 

3,8

 

7,3

 

100

 

N/A

 

12

 

23

 

For transmitters rated at a maximum output power not listed above, the recommended separation distance 

in meters (m) can be estimated 

using the equation applicable to the frequency of the transmitter, where 

is the maximum output power rating of the transmitter in watts (W) 

according to the transmitter manufacturer.   
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies. 
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, 
objects and people.

 

 

Summary of Contents for MFT-9

Page 1: ...re If in doubt about the person s condition seek medical assistance YELLOW Caution Prompt response by the OPERATOR is required GREEN Ready for use Memory Locations There are total 9 sets memories for body temperature Each memory also records the measurement date time mode icon and site location When in power on press the ON MEM button to see the temperature stored with icon F C Switch In power off...

Page 2: ...you hear a beep sound Remove the probe from ear and read the temperature measurement on the LCD After each ear measurement wait the ear icon stop flashing to be ready for next measurement 4 2 Forehead temperature measuring points for attention Remarks a Forehead temperature is displayed in oral mode This mode converts the forehead temperature to display its oral equivalent Value b Before the measu...

Page 3: ...sage Problem Solution Error 5 9 the system is not functioning properly Unload the battery wait for 1 minute and repower it If the message reappears contact the retailer for service Measurement before device stabilization Wait until all the icons stops flashing The ambient temperature is not within the range between 50o F and 104o F 10o C 40o C Allow the thermometer to rest in a room for at least 1...

Page 4: ... domestic purposes Harmonic emissions IEC 61000 3 2 Not applicable Voltage fluctuations flicker emissions IEC 61000 3 3 Not applicable Guidance and manufacturer s declaration electromagnetic immunity The MFT 9 is intended for use in the electromagnetic environment specified below The customer or the user of the MFT 9 should assure that it is used in such an environment Immunity test IEC 60601 test...

Page 5: ... for 5 cycles 70 UT 30 dip in UT for 25 cycles 5 UT 95 dip in UT for 5 sec Not applicable Mains power quality should be that of a typical commercial or hospital environment If the user of the MFT 9 requires continued operation during power mains interruptions it is recommended that the MFT 9 be powered from an uninterruptible power supply or a battery Power frequency 50 60 Hz magnetic field IEC 61...