CareTouch MFT-9 Manual Download Page 2

Figure 4.1.1 

Figure 4.1.2 

Real time clock setting

 

1.

 

Power on: Press the “ON/MEM” button. The thermometer is ready for use after hearing two beep sounds. 

2.

 

Press and hold the “ON/MEM” button for 5sec. When you see the Set icon starts flashing, it indicates that you are in the setting screen.(see the following figure.)   

3.

 

Press the “FOREHEAD” button to make the “24” icon flash. This indicates that you are in the 12/24-hour setting screen. Press the “FOREHEAD” button to set 
12/24-hour Format from “24”(24-hour) to “12”(12-hour). 

4.

 

Press “ON/MEM” button to move the flashing in the sequence shown below to select our settings: 

hour

minutes

year

month

day

leave the setting screen. 

5.

 

While the hour, minutes, year, month, day setting is flashing, use “FOREHEAD” button to change it. 

 

Use of the thermometer

 

Note: The device must stay in stable ambient (room) temperature for 15 minutes before operating. 

1. Always make sure the probe is clean, and without damage.     

 

Warning: Choking from swallowing small parts and batteries by children or pets is possible, 

please keep small parts and batteries at places where children and pets can’t reach.

 

2. Power on: Press “ON/MEM” button.(see figure 1)   

3. Mode selection:   
  3.1

 

Ear mode

: Press the “Ear” button, you can get the body temperature by ear measurement after   

                        hearing two beep sounds. (see figure 1) 

  3.2 Forehead mode

: Press the “FOREHEAD” button, you can get the body temperature by forehead measurement after hearing two beep sounds. (see figure 2) 

  3.3

 

Surface mode

:

 Press and hold the “ON/MEM” button, and press the “FOREHEAD” button one   

                            time

, you can see the    icon on your LCD display(see figure 3). In this mode, 

                    press the “FOREHEAD” button to get the target surface temperature. 

3.4 Stopwatch

:

 Press and hold the “ON/MEM” button, and press the “FOREHEAD” button twice

,   

you can see the    icon on your LCD display(see figure 4). In this mode, press the   
“FOREHEAD” button to count by 0.01sec to 3 minutes.     

4. Temperature taking:   

4.1 Ear temperature measuring, points for attention:   

 

Remarks:

 

a. It is recommended that you measure 3 times with the same ear. If the 3 measurements are different, select the highest temperature. 

b. To avoid the risk of cross contamination, please clean the probe according to “Care and cleaning” section after each use. 
c. The different temperature varies in healthy persons between different parts of the body can be between 32.4~33.8°F (0.2~1°C) 

d. The “Clinical Bias” is 31.1 ~ 30.7°F (-0.5 ~ -0.7°C). 
e. The “Limits of Agreement is 1.02. 

f. The "Repeatability" is 32.3°F (0.19°C). 

 

4.1.1  Gently  pull  the  ear  back  to  straighten  the  ear  canal  (Fig.4.1.1)  and  snugly  position  the  probe  into  the  ear  canal,  aiming 

towards the membrane of the eardrum to obtain an accurate reading (Fig.4.1.2).   

4.1.2  Press  and  hold  the  “Ear”  button  until  you  hear  a  beep  sound.  Remove  the  probe  from  ear  and  read  the  temperature 

measurement on the LCD. After each ear measurement, wait the ear icon stop flashing to be ready for next measurement. 

4.2 Forehead temperature measuring, points for attention:

   

 

Remarks:

 

a. Forehead temperature is displayed in oral mode. This mode converts the forehead temperature to display its “oral-equivalent” Value. 

b. Before the measurement, please stay in a stable environment for 5mins and avoid exercise, bath for 30mins. 

c. Remember to keep the forehead area clean and away from sweat, cosmetics and scar while taking temperature. 

d.

 

The “Clinical Bias” is 29.5 ~ 28.9°F (-1.4 ~ -1.7°C). 

e. The “Limits of Agreement” is 0.98. 
f. The "Repeatability" is 32.4°F (0.20°C). 

 

Figure 5

 

You can hold the thermometer within 1 cm from the central forehead (See Figure 5) and press the "FOREHEAD" button to get the forehead   
measurement. While scanning, you will hear a beep sound, which indicates you the newest measurement is taking place. Measurement has 
been completed after a long beep is heard and the forehead icon stops flashing. The time consuming for measurement might be 1 second. After 
two short beeps are heard, the forehead icon stops flashing to be ready for next measurement.

 

 

4.3 Measuring temperature under surface mode:   

4.3.1 When you press the “FOREHEAD” button, you will get the real time temperature immediately. If you press and hold the “FOREHEAD” button, the reading of 

measurement will be continuous updated.   

4.3.2 Applications include temperature measurements for Water, Milk, Cloth, Skin or other object. 

* Note: This mode shows the actual, unadjusted surface temperatures, which is different from the body temperature. 

4.4 Get the room temperature 

4.4.1 Setting “Real time clock” first to enable this function. After power off, room temperature will be shown on the screen with    icon. 

4.4.2 For room temperature, the thermometer should be placed on the table and avoid the direct sunshine or air conditioner flow on.   
4.4.3 The room temperature can be reference after 15min later. 

5. After measurement: 

5.1 Power off: Device will automatically shut off if left idle for more than 1 minute to extend battery life. The LCD will only show the date, time and room temperature. 
5.2 Clean the probe after each use to ensure an accurate reading and avoid cross contamination.    (See the section of Care and Cleaning for details.) 
5.3 Remember to put on the Probe Cap when not in use.   
 

Use of the Stopwatch

 

1.

 

Press and hold the “ON/MEM” button, and press the “FOREHEAD” button twice for “stopwatch” mode. 

2.

 

For start or stop counting, press the “FOREHEAD” button; for reset counting, press the “ON/MEM” button.   

3.

 

Press and hold the “ON/MEM” button, and press the “FOREHEAD” button to enter the forehead mode. 

4.

 

How to measure the heartbeat? If you count out the number of beating is 18 in 15sec, then you can calculate the heartbeat of patient is 72 times in 1min. 

 

 

 

     

Figure 1        Figure2        Figure 3          Figure 4         

Summary of Contents for MFT-9

Page 1: ...re If in doubt about the person s condition seek medical assistance YELLOW Caution Prompt response by the OPERATOR is required GREEN Ready for use Memory Locations There are total 9 sets memories for body temperature Each memory also records the measurement date time mode icon and site location When in power on press the ON MEM button to see the temperature stored with icon F C Switch In power off...

Page 2: ...you hear a beep sound Remove the probe from ear and read the temperature measurement on the LCD After each ear measurement wait the ear icon stop flashing to be ready for next measurement 4 2 Forehead temperature measuring points for attention Remarks a Forehead temperature is displayed in oral mode This mode converts the forehead temperature to display its oral equivalent Value b Before the measu...

Page 3: ...sage Problem Solution Error 5 9 the system is not functioning properly Unload the battery wait for 1 minute and repower it If the message reappears contact the retailer for service Measurement before device stabilization Wait until all the icons stops flashing The ambient temperature is not within the range between 50o F and 104o F 10o C 40o C Allow the thermometer to rest in a room for at least 1...

Page 4: ... domestic purposes Harmonic emissions IEC 61000 3 2 Not applicable Voltage fluctuations flicker emissions IEC 61000 3 3 Not applicable Guidance and manufacturer s declaration electromagnetic immunity The MFT 9 is intended for use in the electromagnetic environment specified below The customer or the user of the MFT 9 should assure that it is used in such an environment Immunity test IEC 60601 test...

Page 5: ... for 5 cycles 70 UT 30 dip in UT for 25 cycles 5 UT 95 dip in UT for 5 sec Not applicable Mains power quality should be that of a typical commercial or hospital environment If the user of the MFT 9 requires continued operation during power mains interruptions it is recommended that the MFT 9 be powered from an uninterruptible power supply or a battery Power frequency 50 60 Hz magnetic field IEC 61...