CareTouch MFT-9 Manual Download Page 3

Ref No.

072017 

Important Notes 

Care and cleaning

 

  T

he probe is the most delicate part of the thermometer. Use with care when cleaning the lens to avoid damage.   

a. After the measurement, please use the cotton swab with the Alcohol (70% concentration) to clean the inside of the probe, 

including lens and metal parts. 

b. Allow the probe to fully dry for at least 1 minute. 
c.

 

Storage temperature range: It should be stored at room temperature between   

-4~122°F (-20~+50°C

)

, RH

85

%

 

d .Keep the unit dry and away from any liquids and direct sunlight.   
e. The Probe should not be submerged into liquids. 

* Note: Please check the device if damaged once it falls. If you can’t make sure of it ,please send the   
            complete device to the nearest retailer for recalibration.   

  Holding the thermometer too long may cause a higher ambient temperature reading of the probe.   

  This could make the body temperature measurement lower than usual. 

Battery replacement 
 

 

When the “Low Battery” icon indicates the battery is low, the battery should be replaced   
immediately with CR2032 *1pcs battery. 

1

  Open the battery cover: Insert a pointed object, such as a pen, into battery cover pick hole.   

At the same time, use thumb to push battery cover out.   

2

  Hold the device and flip the battery out with a small screwdriver   

3

  Insert the new battery under the  metal hook  on the left side and press the right side of the 

battery down until you hear a click. 

4

  Replace the battery cover 

 

The positive (+) side up and the negative (-) side pointed down.

   

   

 

 

 

 

Trouble shooting: 

Error Message 

Problem 

Solution 

 

Error 5~9, the system is not functioning properly. 

Unload the battery, wait for 1 minute and repower it. If the message 

reappears, contact the retailer for service. 

 

Measurement before device stabilization.

 

Wait until all the icons stops flashing.

 

 

The ambient temperature is not within the range between 50

o

F and 104

o

F (10

o

C ~ 40

o

C). 

Allow the thermometer to rest in a room for at least 15 minutes at 
room temperature: 50

o

F ~104

o

F (10

o

C and 40

o

C). 

 

(1)

 

In ear/forehead mode: Temperature taken is higher than +108°F (42.2

o

C) 

(2)

 

In surface mode: Temperature taken is higher than +176°F (80

o

C) 

Please select the target within specifications. If a malfunction still 

exists, please contact the nearest retailer. 

 

(1)

 

In ear/forehead mode: Temperature taken is lower than +93.2°F (34

o

C) 

(2)

 

In surface mode: Temperature taken is lower than –7.6°F (22

o

C) 

 

Device cannot be powered on to the ready stage. 

Change with a new battery. 

 

Warranty:

 

Warranty: 12 months   
Manufacture Date: as the serial number (please open the battery cover, it is shown on the inside of the device.)

   

 

Ex.SN:E512A000001, the first “E” is External, the second number “5” is the last number of manufacture year, the third and the fourth number “12” is the manufacture 
month, the others is the serial number.

 

Note: The thermometer is calibrated at the time of manufacture. If at any time you question the accuracy of temperature measurements, please contact the 

dealers or nearest service address. 

 

 

Warning: No modification of this equipment is allowed.

 

 

Please read the instructions for use       

  BF type applied part 

 

Manufactured in China for Future Diagnostics LLC 
266 47th Street, Brooklyn, NY11220 
Package Design © 2017 
CareTouch is a Trademark of Future Diagnostics LLC

 

 

 

Summary of Contents for MFT-9

Page 1: ...re If in doubt about the person s condition seek medical assistance YELLOW Caution Prompt response by the OPERATOR is required GREEN Ready for use Memory Locations There are total 9 sets memories for body temperature Each memory also records the measurement date time mode icon and site location When in power on press the ON MEM button to see the temperature stored with icon F C Switch In power off...

Page 2: ...you hear a beep sound Remove the probe from ear and read the temperature measurement on the LCD After each ear measurement wait the ear icon stop flashing to be ready for next measurement 4 2 Forehead temperature measuring points for attention Remarks a Forehead temperature is displayed in oral mode This mode converts the forehead temperature to display its oral equivalent Value b Before the measu...

Page 3: ...sage Problem Solution Error 5 9 the system is not functioning properly Unload the battery wait for 1 minute and repower it If the message reappears contact the retailer for service Measurement before device stabilization Wait until all the icons stops flashing The ambient temperature is not within the range between 50o F and 104o F 10o C 40o C Allow the thermometer to rest in a room for at least 1...

Page 4: ... domestic purposes Harmonic emissions IEC 61000 3 2 Not applicable Voltage fluctuations flicker emissions IEC 61000 3 3 Not applicable Guidance and manufacturer s declaration electromagnetic immunity The MFT 9 is intended for use in the electromagnetic environment specified below The customer or the user of the MFT 9 should assure that it is used in such an environment Immunity test IEC 60601 test...

Page 5: ... for 5 cycles 70 UT 30 dip in UT for 25 cycles 5 UT 95 dip in UT for 5 sec Not applicable Mains power quality should be that of a typical commercial or hospital environment If the user of the MFT 9 requires continued operation during power mains interruptions it is recommended that the MFT 9 be powered from an uninterruptible power supply or a battery Power frequency 50 60 Hz magnetic field IEC 61...

Reviews: