CareTouch MFT-9 Manual Download Page 1

  MFT-9 Multi-function Infrared Thermometer 

 

Functions 

Real Time Clock 

The real time clock will be recorded with the memory function and help you to recognize each 
measurement result. (It includes the 12/24 Format_Hour_Minutes_Date.)   

 Please see the “Real time clock setting” section to learn how to setup the time in the first use. 

 

 

 

Room Temperature

 

Suitable  ambient  temperature  is  important  for  the  baby  and  patient,  the  thermometer  always  help  you 
recognize the room temperature. 

 Setting “Real time clock” first to enable this function. After power off, room temperature will be shown   

on the screen with    icon. 

 Please see the “Use of the thermometer” section to learn how to get the correct room temperature.   

 

Ear / Forehead 
Temperature 

The  thermometer  has  been  designed  for  practical  use.  It’s  not  meant  to  replace  a  visit  to  the  doctor. 
Please  also  remember  to  compare  the  measurement  result  to  your  regular  body  temperature.

  Please 

consult with doctor if you have health concerns.

 

 Please see the “Use of the thermometer” section to learn how to measure the body temperature. 

 

 

Surface Mode 

The surface mode shows the actual, unadjusted surface temperatures,  which is different from the body 
temperature. It can help you  to  monitor if the  object temperature is suitable for the  baby or patient, for 
example the baby’s milk. 

 Please see the “Use of the thermometer” section to learn how to measure the object temperature. 

 

Stopwatch 

The thermometer built the stopwatch function to help you to monitor the heartbeat since heartbeat is one 
of the healthy indexes for people.   

 Please see the “Use of the Stopwatch” section to learn how to measure the heartbeat result. 

 

Temperature Alarm 

If thermometer detects a body temperature 

99.5°F (37.5°C), there will be a long beep sound followed 

by three short beep sound to warn the user for the temperature is above 99.5°F (37.5°C). 

 

Zone Indicator 

Under ear or forehead mode, if thermometer detects a body temperature <99.5°F (37.5°C), there will be " 
Green" light signal; if temperature 

99.5°F (37.5°C) and <100.4°F (38.0°C), there will be" Yellow" light 

signal; if temperature 

100.4°F (38.0°C), there will be" Red" light signal. 

  RED: Warning - Immediate response by the OPERATOR is required. Suggest continually monitoring   

the body temperature. If in doubt about the person’s condition, seek medical assistance. 

YELLOW: Caution - Prompt response by the OPERATOR is required. 

GREEN: Ready for use.

 

 

Memory Locations

 

There  are  total  9  sets  memories  for  body  temperature.  Each  memory  also  records  the  measurement 
date/time/mode icon and site location. 

 When in power on, press the “ON/MEM” button to see the temperature stored with 

  icon. 

 

°F / °C Switch

 

 In “power off” mode (The LCD only shows the date, time and room temperature). Press and hold the 

“FOREHEAD”  button,  then  immediately  press  and  release  the  “ON/MEM”  button.  After  3  seconds,  icon 
“°F” will be switched to icon “°C”. (You may use the same way to switch back to “°F”) 

 

 

 

 

 

Intended use: 

The infrared ear/forehead thermometer can measure the infrared heat generated 
by eardrum/central forehead area and surrounding tissues to reflect patient’s 
body temperature accurately. 

 
 
 

Intended operator:  At least 11 years old (5 years intensive reading experience), no maximum. 

 

Specifications 

 

Temperature measurement range:   

Ear / Forehead mode: 93.2~108°F (34~42.2°C), Surface mode: -7.6~176°F (-22~80°C) 

 

Operating temperature range: 50~104°F (10~40°C), 15%~85% RH 

 

Storage temperature range: It should be stored at room temperature between -4~122°F 
(-20~+50°C), RH

85% 

Transportation temperature shall be less than 158°F (70°C), RH

95% 

 

Atmospheric pressure: 800~1013 hPa 

 

Comply with ASTM E1965-98, EN ISO 80601-2-56, IEC/EN60601-1-2(EMC), 

IEC/EN60601-1(Safety) standards, ISO10993, RoHS. 

 

Accuracy: 

    For ear / forehead mode: 

±

0.4°F (0.2°C) within 95~107.6°F (35~42°C),   

                                         

±

0.5°F (0.3°C) for other range. 

For surface mode: 

±

0.5°F (0.3°C) within 71.6~108°F (22~42.2°C),   

                            others 

±

4% or 

±

4°F (2°C) whichever is greater. 

 

Temperature alarm & Memory locations under “Ear/Forehead mode”   

(with each measurement date/time/mode icon) 

 

3 color LED Zone Indicator 

 

Real time clock, °F / °C Switch mode 

 

Battery: one lithium cell battery (CR2032 *1pcs) 

 

Battery life: around 3,000

 

continuous readings.   

 

Expected Service Life: 4 years 

 

This thermometer converts the ear/forehead temperature to display its “oral equivalent.” 
(according to the result of the clinical evaluation to get the offset value) 

 

Enclosure Rating: IP22 

 

Dimensions: 146.4 x 38.3 x 54.8 mm 

 

Weight: 63.3 grams including battery

 

 

The device should not submerge into any liquids and expose it to direct moisture.

 

 

There is no gender and age limitation for using infrared thermometer.   

  This is not an AP

 

or APG product.

 

 
 
 
 

 
 
 
 

            Before the measurement, please make sure the selected 

 

button accords with the measured position. 

 

Summary of Contents for MFT-9

Page 1: ...re If in doubt about the person s condition seek medical assistance YELLOW Caution Prompt response by the OPERATOR is required GREEN Ready for use Memory Locations There are total 9 sets memories for body temperature Each memory also records the measurement date time mode icon and site location When in power on press the ON MEM button to see the temperature stored with icon F C Switch In power off...

Page 2: ...you hear a beep sound Remove the probe from ear and read the temperature measurement on the LCD After each ear measurement wait the ear icon stop flashing to be ready for next measurement 4 2 Forehead temperature measuring points for attention Remarks a Forehead temperature is displayed in oral mode This mode converts the forehead temperature to display its oral equivalent Value b Before the measu...

Page 3: ...sage Problem Solution Error 5 9 the system is not functioning properly Unload the battery wait for 1 minute and repower it If the message reappears contact the retailer for service Measurement before device stabilization Wait until all the icons stops flashing The ambient temperature is not within the range between 50o F and 104o F 10o C 40o C Allow the thermometer to rest in a room for at least 1...

Page 4: ... domestic purposes Harmonic emissions IEC 61000 3 2 Not applicable Voltage fluctuations flicker emissions IEC 61000 3 3 Not applicable Guidance and manufacturer s declaration electromagnetic immunity The MFT 9 is intended for use in the electromagnetic environment specified below The customer or the user of the MFT 9 should assure that it is used in such an environment Immunity test IEC 60601 test...

Page 5: ... for 5 cycles 70 UT 30 dip in UT for 25 cycles 5 UT 95 dip in UT for 5 sec Not applicable Mains power quality should be that of a typical commercial or hospital environment If the user of the MFT 9 requires continued operation during power mains interruptions it is recommended that the MFT 9 be powered from an uninterruptible power supply or a battery Power frequency 50 60 Hz magnetic field IEC 61...

Reviews: