Safety and Regulatory Information
1–20
AJ4309 | 2019-08-13
Proximity Fields From RF Wireless Communications Equipment
Input AC Power Port
Test Frequency
(MHz)
Band (MHz)
Test Levels
Professional healthcare facility environment
385
380–390
Pulse modulation 18 Hz, 27 V/m
450
430–470
FM, ±5kHz deviation, 1 kHz sine, 28 V/m
710
704–787
Pulse modulation 217 Hz, 9 V/m
745
780
810
800-–960
Pulse modulation 18 Hz, 28 V/m
870
930
1720
1700–1990
Pulse modulation 217 Hz, 28 V/m
1845
1970
2450
2400–2570
Pulse modulation 217 Hz, 28 V/m
5240
5100–5800
Pulse modulation 217 Hz, 9 V/m
5500
5785
Emissions Test
EMC Standard
Test Levels
Professional healthcare facility environment
Electrical fast transients/burst
IEC 61000-4-4
±2 kV
100 kHz repetition frequency
Surges
Line-to-line
IEC 61000-4-5
±0.5 kV, ±1kV