Infant Flow
®
LP SiPAP Model M675
675-120 Rev. G
Alarm
Method of Setting
Indications and
Actions
Method of Resetting
or Canceling
Low oxygen concentration
(< 5 FiO
2
below set point
at time alarm is set for 15
seconds,
or
< 20 FiO
2
for > 15
seconds).
Set automatically on
entering NCPAP
mode for the first time
after start up or by
pushing the alarm
button for three
seconds at any time.
Intermittent high level
Audible Alarm.
Warning Bar flashes.
O
2
digital display
flashes.
Digital O
2
alarm low
limit is highlighted.
Alarm Button flashes.
Push the Alarm
Button once to stop
the Audible Alarm for
30 seconds (Alarm
Button flashes, digital
O
2
alarm low limit
stays highlighted,
Warning Bar still
flashes).
Push the Alarm
Button for three
seconds to reset the
limit (alarms clear).
High NCPAP pressure (>
3 cmH
2
O above set point
at time alarm is set for 15
seconds).
Set automatically on
entering NCPAP
mode for the first time
after start up or by
pushing the alarm
button for three
seconds at any time.
Intermittent high level
Audible Alarm.
Warning Bar flashes.
Pressure display
flashes.
Digital pressure high
limit is highlighted.
Alarm Button flashes.
Push the Alarm
Button once to stop
the audible alarm for
30 seconds (Alarm
Button flashes, digital
pressure high limit
stays highlighted,
Warning Bar still
flashes). Push the
Alarm Button for
three seconds to
reset the limit (alarms
clear).
Low pressure
(< 2 cmH
2
O below set
point for 15 seconds)
or
< 1 cmH
2
O at any time).
Set automatically on
entering NCPAP
mode for the first time
after start up or by
pushing the alarm
button for three
seconds at any time.
Intermittent high level
Audible Alarm.
Warning Bar flashes.
Pressure display
flashes.
Digital pressure low
limit is highlighted.
Alarm Button flashes.
Push the Alarm
Button once to stop
the Audible Alarm for
30 seconds (Alarm
Button flashes, digital
pressure low limit
stays highlighted,
Warning Bar still
flashes ).
Push the Alarm
Button for three
seconds to reset the
limit (alarms clear).
Summary of Contents for Infant Flow LP SiPAP
Page 1: ...Infant Flow LP SiPAP and SiPAP Model M675 Service manual ...
Page 16: ...6 Infant Flow LP SiPAP Model M675 675 120 Rev G Infant Flow LP SiPAP and SiPAP Model M675 ...
Page 20: ...10 Infant Flow LP SiPAP Model M675 675 120 Rev G Infant Flow LP SiPAP and SiPAP Model M675 ...
Page 24: ...14 Infant Flow LP SiPAP Model M675 675 120 Rev G Figure 3 IFSD Internal Components ...
Page 30: ...20 Infant Flow LP SiPAP Model M675 675 120 Rev G Figure 6 IFSD Gas Flow Schematic ...
Page 88: ...78 Infant Flow LP SiPAP Model M675 675 120 Rev G Infant Flow LP SiPAP and SiPAP Model M675 ...
Page 90: ...80 Infant Flow LP SiPAP Model M675 675 120 Rev G Infant Flow LP SiPAP and SiPAP Model M675 ...
Page 93: ...Service Manual 83 675 120 Rev G Appendix D Pneumatics Assembly ...
Page 94: ...84 Infant Flow LP SiPAP Model M675 675 120 Rev G ...