Table 5 - Faults and Indications
Alarm
Method of Setting
Indications and
Actions
Method of Resetting
or Canceling
Minimum oxygen
concentration
(< 18 FiO
2
).
Always active when
power is on.
Intermittent high level
Audible Alarm.
Warning Bar flashes.
FiO
2
display flashes.
Digital O
2
alarm low
limit is highlighted.
Alarm Button flashes.
Restore FiO
2
level to
above the low limit
then push Alarm
Button for three
seconds.
Over pressure
(Patient pressure
> 11 cmH
2
O when in
NCPAP mode).
Always active when
power is on.
Intermittent high level
Audible Alarm.
Warning Bar flashes.
Pressure display
flashes.
Digital high-pressure
alarm limit is
highlighted.
Alarm Button flashes.
Dump valve actuated
for three seconds to
stop flow to patient and
repeated until flow
returns to normal.
Reduce pressure to
below the high
pressure limit then
push the Alarm
Button for three
seconds.
High oxygen concentration
(> 5 FiO
2
above set point
at time alarm is set for 15
seconds,
or
> 104 FiO
2
for > 15
seconds).
Set automatically on
entering NCPAP
mode for the first time
after start up or by
pushing the Alarm
Button for three
seconds at any time.
Intermittent high level
Audible Alarm.
Warning Bar flashes.
O
2
digital display
flashes.
Digital O
2
alarm high
limit is highlighted.
Alarm Button flashes.
Push the Alarm
Button once to stop
the Audible Alarm for
30 seconds (Alarm
Button flashes, digital
O
2
alarm high limit
stays highlighted,
Warning Bar still
flashes).
Push the Alarm
Button for three
seconds to reset the
limit (alarms clear).
Summary of Contents for Infant Flow LP SiPAP
Page 1: ...Infant Flow LP SiPAP and SiPAP Model M675 Service manual ...
Page 16: ...6 Infant Flow LP SiPAP Model M675 675 120 Rev G Infant Flow LP SiPAP and SiPAP Model M675 ...
Page 20: ...10 Infant Flow LP SiPAP Model M675 675 120 Rev G Infant Flow LP SiPAP and SiPAP Model M675 ...
Page 24: ...14 Infant Flow LP SiPAP Model M675 675 120 Rev G Figure 3 IFSD Internal Components ...
Page 30: ...20 Infant Flow LP SiPAP Model M675 675 120 Rev G Figure 6 IFSD Gas Flow Schematic ...
Page 88: ...78 Infant Flow LP SiPAP Model M675 675 120 Rev G Infant Flow LP SiPAP and SiPAP Model M675 ...
Page 90: ...80 Infant Flow LP SiPAP Model M675 675 120 Rev G Infant Flow LP SiPAP and SiPAP Model M675 ...
Page 93: ...Service Manual 83 675 120 Rev G Appendix D Pneumatics Assembly ...
Page 94: ...84 Infant Flow LP SiPAP Model M675 675 120 Rev G ...