24
Infant Flow
®
LP SiPAP Model M675
675-120 Rev. G
Fault Management
When a software detectable fault condition occurs, the unit still allows a basic level of
treatment to the patient. Table 4 shows the fault conditions and the level of control
available.
Table 4 - Faults, Available Modes and Control Functions
Measurements
Software Control Modes
Alarm
Functionality
FiO
2
CmH
2
O
NCPAP
NCPAP+
Apnea
BiPhasic
+ Apnea
BiPhasic
Tr
Fault lockout
screen shows
error
code(s).
Unit inoperable by
software, operable
in
unpowered
ti
X
X
X
X
X
X
Error code(s)
shown
on mode
selection
screen.
Status bar
mode
alternates
Unit restricted to
NCPAP and
NCPAP+Apnea
modes.
X
X
Error code(s)
shown
on mode
selection
screen.
Status bar
mode
lt
t
Patient trigger
(NCPAP+Apnea,
BApnea
and
BiPhasic Tr)
modes not
available.
X
X
Battery status
shows
discharged
battery.
Unit will not
operate
on battery power
when external
power
supply is
removed.
Summary of Contents for Infant Flow LP SiPAP
Page 1: ...Infant Flow LP SiPAP and SiPAP Model M675 Service manual ...
Page 16: ...6 Infant Flow LP SiPAP Model M675 675 120 Rev G Infant Flow LP SiPAP and SiPAP Model M675 ...
Page 20: ...10 Infant Flow LP SiPAP Model M675 675 120 Rev G Infant Flow LP SiPAP and SiPAP Model M675 ...
Page 24: ...14 Infant Flow LP SiPAP Model M675 675 120 Rev G Figure 3 IFSD Internal Components ...
Page 30: ...20 Infant Flow LP SiPAP Model M675 675 120 Rev G Figure 6 IFSD Gas Flow Schematic ...
Page 88: ...78 Infant Flow LP SiPAP Model M675 675 120 Rev G Infant Flow LP SiPAP and SiPAP Model M675 ...
Page 90: ...80 Infant Flow LP SiPAP Model M675 675 120 Rev G Infant Flow LP SiPAP and SiPAP Model M675 ...
Page 93: ...Service Manual 83 675 120 Rev G Appendix D Pneumatics Assembly ...
Page 94: ...84 Infant Flow LP SiPAP Model M675 675 120 Rev G ...