Features of the Workstation

System Fault Indication

Continuous monitoring of the power distribution and communications system integrity is performed by the Workstation . In the event 

that a system fault occurs whilst in use, the System Fault Indicator will be illuminated accompanied by an audible tone . To avoid any 

possible interruption to the infusion, AC power to the Pumps will be maintained on the MDI tile should a system fault be detected . The 

Workstation briefly illuminates the System Fault Indicator and activates the audible tone each time the Workstation is switched on .

Caution:  If the System Fault Indicator fails to illuminate when the Workstation is switched on, remove the Workstation from 

service and contact Qualified Service Personnel.

Caution:  Should a System Fault occur during use, remove the Workstation from service as soon as practical and contact 

Qualified Service Personnel.

Pump Alarm Location Beacon (where fitted)

The Pump Alarm Location Beacon is mounted on the upper face of the Workstation to assist with identifying the location of any Pumps 

that have entered an alarm or warning state . When lit, the Pump Alarm Location Beacon colour matches that of the visual status indicator 

on the Pumps; red for alarms and amber for warnings . Alarms of high priority, such as when an infusion is terminated are reflected 

as Red . Alarms or warnings of lower priority, such as Near-end-of-Infusion, are reflected as Amber . The Pump Alarm Location Beacon 

will indicate the highest level of alarm if there are multiple pumps in alarm . The Pump Alarm Location Beacon flashes automatically 

whenever any Pump located on an MDI Tile of the Workstation enters the alarm or warning condition, and stops when the condition is 

cleared on the Pump . The Pump Alarm Location Beacon automatically illuminates red then amber each time the Workstation is switched 

on .
The Pump Alarm Location Beacon is provided in order to allow Pumps with active alarms or warnings to be easily located; it does not 

replace the alarm or warning on the Pump which remains the principle indicator that the attention of a clinician is required .

Caution:  If the Pump Alarm Location Beacon fails to illuminate when the Workstation is switched on, suspect a fault with the 

beacon. Remove the Workstation from service and contact Qualified Service Personnel.

A

The Pump Alarm Location Beacon is selectively populated and cannot be fitted to a 4 MDI Tile Horizontal Module. 

However the Pumps have their own means to alarm and it is the hospital’s responsibility to monitor.

A

For the Alaris GW Volumetric Pump, which does not have a visual status indicator: 

•  Alarm State - will activate the Pump Alarm Location Beacon of the Workstation 
•  Warning State - will not activate the Pump Alarm Location Beacon of the Workstation 

Note:  See the ‘Introduction’ section for lists of the Pump Alarm Location Beacon and Nurse Call compatibility between the 

Pumps and Workstation.

Nurse Call Interface

A nurse call interface is provided which is activated whenever any Pump attached to the Workstation enters an alarm or warning state, 

and stops when the condition is cleared on the Pump . The Nurse Call Interface may be connected to operate in a normally open or 

normally closed contact position . Verify the Nurse Call is automatically activated each time the Workstation is switched on .
The Nurse Call interface on the Workstation allows a single connection point to a hospital Nurse Call system . This Nurse Call activates 

when a Pump communicates an alarm or warning condition to the Workstation via the IrDA interface or RS232 for Alaris SE Pump . The 

communication of such an alarm or warning condition could be interrupted if, for example, the Pump was not correctly attached to the 

Workstation .

Caution:  If the Nurse Call Interface fails to operate when the Workstation is switched on, suspect a fault with the interface. 

Remove the Workstation from service and contact Qualified Service Personnel.

Caution:  There will be a delay, of up to 5 seconds, between the Pump going into alarm and the Nurse Call Interface activating 

via the Workstation.

Caution:  Where reliability of the nurse call interface is paramount, then connection has to be made from the nurse call system 

directly to the nurse call interface on the Pump.

Summary of Contents for Alaris Gateway Workstation

Page 1: ...Alaris Gateway Workstation v1 2 Directions For Use en...

Page 2: ...to Infusion Pumps 14 Powering an auxiliary Workstation 14 System Fault Indication 15 Pump Alarm Location Beacon where fitted 15 Nurse Call Interface 15 Operating Precautions 16 Operating Environment 1...

Page 3: ...24 Wireless Ethernet Interface where fitted 24 Auxiliary Interface 24 Product Specifications 25 Electrical 25 Battery 25 Environmental 25 Classification 25 Regulatory Compliance 25 Potential Equalisa...

Page 4: ...2 SN T 135070253 17 04 2014 This is an entry level Workstation with organization and central AC power supply functionality It does not support PDMS connectivity It may be used as an Auxiliary Workstat...

Page 5: ...Communication Engine The Alaris DS Docking Station can be identified by the number of vertical MDI tiles as it has an even number compared to the Workstation which has an odd number and that the AC p...

Page 6: ...es Alaris GH Guardrails Syringe Pump with Plus Software MK3 8002MED01 G v4 1 4 Yes Yes Yes Yes Alaris CC Guardrails Syringe Pump with Plus Software MK3 8003MED01 G v4 1 4 Yes Yes Yes Yes Alaris GH Syr...

Page 7: ...con and nurse call for medium priority alarms however will activate the Pump Alarm Location Beacon and nurse call for high priority alarms To access and configure the software installed on the Worksta...

Page 8: ...0203UNSxx where xx indicates the connectivity option The two or three numbers at the end of the SKU identifies the Workstation configuration see below for all available configurations Available Config...

Page 9: ...ed by contacting your local CareFusion representative Conventions used in this manual BOLD Used for Display names software commands controls and indicators referenced in this manual for example Batter...

Page 10: ...illuminated the Workstation is operating from internal battery when flashing the battery power is low and auto power down is imminent k k AC Power Indicator When illuminated the Workstation is connec...

Page 11: ...lt accompanying documents x Potential Equalisation PE Connector IPX1 Protected against vertically falling drops of water r Alternating Current u Workstation complies with the requirements of Council D...

Page 12: ...ng Diode LED Adjustable Height Pole Slip Catch Pump Alarm Location Beacon Not fitted on 4 MDI Tile Horizontal Module Bag Supports w Covered under Patents U S Pat 6 593 528 EP 1502612 DE CH FR GB ES IT...

Page 13: ...ion Configuration 80203UNSxx 235 shown mounted on an Alaris Trolley Infusion Line Tidies Bag Supports Trolley Bag Support Rail Mounting Rail Mounting Rail Mounting Rail Note Qualified Service Personne...

Page 14: ...where fitted The 18 mm diameter adjustable height pole has been designed as a convenient means of securing the fluid bags onto the Workstation The pole supports a maximum load to 3 kg The pole is hel...

Page 15: ...of a power loss the Workstation with fully charged battery will emit an audible tone every 30 seconds for the first 14 minutes to alert operators This audible tone will escalate to every 15 seconds af...

Page 16: ...the alarm or warning on the Pump which remains the principle indicator that the attention of a clinician is required Caution If the Pump Alarm Location Beacon fails to illuminate when the Workstation...

Page 17: ...IEC EN60601 1 2 If however the Workstation interacts with other equipment measures should be taken to minimise the effects for instance by repositioning or relocation Therapeutic Radiation Equipment...

Page 18: ...moisture fluid spillage high humidity high temperature suspicion of damage When transporting or storing the Workstation use original packaging where possible and adhere to temperature humidity and pr...

Page 19: ...ndle is in full working order before use Warning The pole clamp should be used to mount the Workstation on fixed poles overhead swing arm poles only Warning People of a weak disposition should not att...

Page 20: ...d B Status indicators will flash after a long beep Not applicable 12 System initiates the power on Self test Verify the System Fault indicator illuminates briefly and is then extinguished Not applicab...

Page 21: ...hen the Pump is securely attached it should not come off the Workstation Rectangular bar Rotating cam Release lever push to release When using volumetric pumps with the Workstation it is recommended t...

Page 22: ...be used on a 10 Base T 100 Base Tx switched LAN A DHCP client service allows the use of either fixed or dynamic network addressing of the Workstation Similarly a DNS client is provided Configuration...

Page 23: ...tatus of the reader All LEDs off The barcode reader is not receiving power or the barcode reader is in standby mode Steady Green The laser is active Steady Green Single Red Flash The barcode has been...

Page 24: ...ting 30V 1A Description Pin 1 NC_COM 1 2 3 Pin 2 NC_NC Pin 3 NC_NO Barcode Reader Interface For use with CareFusion supplied barcode reader only Workstation Connector Type Binder 09 0998 00 05 Cable T...

Page 25: ...ed in the Workstation is in compliance with the essential requirements and other relevant provisions of Radio and Telecom Terminal Equipment Directive 1999 5 EC Auxiliary Interface Connector Type RJ45...

Page 26: ...nd an 11 pump load Operating time 20 minutes Environmental Operating Transport and Storage Temperature 5 C 40 C 20 C 50 C Humidity 20 90 15 95 Atmospheric Pressure 700 1060hPa 500 1060hPa Classificati...

Page 27: ...13 39 320 Yes 80203UNSxx 235 1240 485 155 14 5 45 5 360 Yes Trolley Compatibility Configuration1 Trolley Compatible 80083UN00 00 5 Alaris Trolley Compatible 80203UNS00 00 6 80203UNSxx 30 80203UNSxx 3...

Page 28: ...unit leakage current must be measured If more than 500 A the equipment should not be used but should be serviced by Qualified Service Personnel Battery It is recommended that only Qualified Service Pe...

Page 29: ...s fluid build up on theWorkstation Do not use aggressive cleaning agents as these may damage the exterior surface of the Workstation Do not steam autoclave ethylene oxide sterilise or immerse this Wor...

Page 30: ...and password are required to access our manuals Please contact a local customer services representative to obtain login details Part Number Description 0000EL00889 AC Power Lead UK 0000EL00938 AC Pow...

Page 31: ...en 8 10 3994 DB Houten Nederland CareFusion 10020 Pacific Mesa Blvd San Diego CA 92121 USA Tel 32 0 2 267 38 99 Tel 34 902 555 660 Tel 31 0 30 2289 711 Tel 1 800 854 7128 Fax 32 0 2 267 99 21 Fax 34 9...

Page 32: ...of CareFusion Corporation or one of its affiliates and its receipt or possession does not convey any rights to reproduce its contents or to manufacture or sell any product described Reproduction discl...

Reviews: