CareFusion Alaris Gateway Workstation Directions For Use Manual Download Page 15

1000DF00742 Issue 2

14/30

Alaris™ Gateway Workstation v1.2

Features of the Workstation

Power Input

The  Workstation is powered from the AC supply through a standard IEC AC connector . When connected to the AC supply the AC Power 

indicator is illuminated . Both the Live and Neutral lines of the main supply are protected using fuses carried in a double fuse holder 

located on the AC inlet connector .

Warning:   When connected to the AC supply, a three wire (Live, Neutral, Earth) supply must be used. If the integrity of the 

external protective conductor in the installation or its arrangement is in doubt, then the Workstation must not be 

used.

A

To isolate the Workstation from AC supply remove the AC connector from the source socket. The Workstation should 

be positioned to allow access for disconnecting the AC connector.

Battery Supply

The Workstation should normally be operated from the AC power supply . However, in the event of temporary loss of AC power, an 

internal power supply will provide approximately 20 minutes of communications only . AC power to the Pumps will be discontinued .

Warning:   AC Power should be re-applied as soon as possible as there will be no AC power supplied to the Pumps while the 

Workstation is running on the battery. 

The Battery indicator illuminates whenever the Workstation is running from the internal battery . When illuminated the Workstation is 

operating from internal battery; when flashing the battery power is low and auto power down is imminent . The battery is automatically 

charged whenever the Workstation is connected to the AC supply . As the Workstation is designed to operate from the AC power supply it 

will only power up when connected to the AC supply .

Caution:   In the event of a power loss, the Workstation with fully charged battery will emit an audible tone every 30 seconds 

for the first 14 minutes to alert  operators. This audible tone will escalate to every 15 seconds after 14 minutes until 

the battery is fully depleted.

A

If transfer of the Workstation is required, then prior to disconnecting from AC power supply the User must ensure 

there is sufficient battery power on each of the Pumps.

AC Power Output to Infusion Pumps

The Workstation has its own power distribution circuit to supply AC power to the attached Pumps . For safety, power is not applied to the 

MDI tile IEC connector until the Pump is fully attached to the MDI tile . The AC power indicator on the infusion Pump will illuminate when 

Pump is powered and charging .

Warning:  The MDI tile AC outlet connection is intended only for connection to a Pump. Never attach any other equipment to 

the outlet connector.

The Workstation minimises the potential for a high peak in-rush currents when AC power is simultaneously applied to the Pumps . When 

the Workstation is initially switched on, or when re-connected to the AC whilst operating from the internal battery, a small delay in the 

application of AC power will occur between each MDI tile . This staggers the distribution of AC power to all Pumps and therefore, reduces 

the peak in-rush current .

Powering an auxiliary Workstation

Workstations are fitted with an auxiliary AC power outlet connection . However, this should not be used to power a second Workstation . If 

a second Workstation is required at the bedside, plug the Auxiliary Workstation directly into a wall AC power outlet .

Note:

  The safe use of the auxiliary AC power outlet requires management of the 

number of Pumps installed so that the system earth leakage current does not 

exceed 500uA . For this reason a warning label is placed blocking access to 

auxiliary AC outlet . This warning label should not be removed .

Summary of Contents for Alaris Gateway Workstation

Page 1: ...Alaris Gateway Workstation v1 2 Directions For Use en...

Page 2: ...to Infusion Pumps 14 Powering an auxiliary Workstation 14 System Fault Indication 15 Pump Alarm Location Beacon where fitted 15 Nurse Call Interface 15 Operating Precautions 16 Operating Environment 1...

Page 3: ...24 Wireless Ethernet Interface where fitted 24 Auxiliary Interface 24 Product Specifications 25 Electrical 25 Battery 25 Environmental 25 Classification 25 Regulatory Compliance 25 Potential Equalisa...

Page 4: ...2 SN T 135070253 17 04 2014 This is an entry level Workstation with organization and central AC power supply functionality It does not support PDMS connectivity It may be used as an Auxiliary Workstat...

Page 5: ...Communication Engine The Alaris DS Docking Station can be identified by the number of vertical MDI tiles as it has an even number compared to the Workstation which has an odd number and that the AC p...

Page 6: ...es Alaris GH Guardrails Syringe Pump with Plus Software MK3 8002MED01 G v4 1 4 Yes Yes Yes Yes Alaris CC Guardrails Syringe Pump with Plus Software MK3 8003MED01 G v4 1 4 Yes Yes Yes Yes Alaris GH Syr...

Page 7: ...con and nurse call for medium priority alarms however will activate the Pump Alarm Location Beacon and nurse call for high priority alarms To access and configure the software installed on the Worksta...

Page 8: ...0203UNSxx where xx indicates the connectivity option The two or three numbers at the end of the SKU identifies the Workstation configuration see below for all available configurations Available Config...

Page 9: ...ed by contacting your local CareFusion representative Conventions used in this manual BOLD Used for Display names software commands controls and indicators referenced in this manual for example Batter...

Page 10: ...illuminated the Workstation is operating from internal battery when flashing the battery power is low and auto power down is imminent k k AC Power Indicator When illuminated the Workstation is connec...

Page 11: ...lt accompanying documents x Potential Equalisation PE Connector IPX1 Protected against vertically falling drops of water r Alternating Current u Workstation complies with the requirements of Council D...

Page 12: ...ng Diode LED Adjustable Height Pole Slip Catch Pump Alarm Location Beacon Not fitted on 4 MDI Tile Horizontal Module Bag Supports w Covered under Patents U S Pat 6 593 528 EP 1502612 DE CH FR GB ES IT...

Page 13: ...ion Configuration 80203UNSxx 235 shown mounted on an Alaris Trolley Infusion Line Tidies Bag Supports Trolley Bag Support Rail Mounting Rail Mounting Rail Mounting Rail Note Qualified Service Personne...

Page 14: ...where fitted The 18 mm diameter adjustable height pole has been designed as a convenient means of securing the fluid bags onto the Workstation The pole supports a maximum load to 3 kg The pole is hel...

Page 15: ...of a power loss the Workstation with fully charged battery will emit an audible tone every 30 seconds for the first 14 minutes to alert operators This audible tone will escalate to every 15 seconds af...

Page 16: ...the alarm or warning on the Pump which remains the principle indicator that the attention of a clinician is required Caution If the Pump Alarm Location Beacon fails to illuminate when the Workstation...

Page 17: ...IEC EN60601 1 2 If however the Workstation interacts with other equipment measures should be taken to minimise the effects for instance by repositioning or relocation Therapeutic Radiation Equipment...

Page 18: ...moisture fluid spillage high humidity high temperature suspicion of damage When transporting or storing the Workstation use original packaging where possible and adhere to temperature humidity and pr...

Page 19: ...ndle is in full working order before use Warning The pole clamp should be used to mount the Workstation on fixed poles overhead swing arm poles only Warning People of a weak disposition should not att...

Page 20: ...d B Status indicators will flash after a long beep Not applicable 12 System initiates the power on Self test Verify the System Fault indicator illuminates briefly and is then extinguished Not applicab...

Page 21: ...hen the Pump is securely attached it should not come off the Workstation Rectangular bar Rotating cam Release lever push to release When using volumetric pumps with the Workstation it is recommended t...

Page 22: ...be used on a 10 Base T 100 Base Tx switched LAN A DHCP client service allows the use of either fixed or dynamic network addressing of the Workstation Similarly a DNS client is provided Configuration...

Page 23: ...tatus of the reader All LEDs off The barcode reader is not receiving power or the barcode reader is in standby mode Steady Green The laser is active Steady Green Single Red Flash The barcode has been...

Page 24: ...ting 30V 1A Description Pin 1 NC_COM 1 2 3 Pin 2 NC_NC Pin 3 NC_NO Barcode Reader Interface For use with CareFusion supplied barcode reader only Workstation Connector Type Binder 09 0998 00 05 Cable T...

Page 25: ...ed in the Workstation is in compliance with the essential requirements and other relevant provisions of Radio and Telecom Terminal Equipment Directive 1999 5 EC Auxiliary Interface Connector Type RJ45...

Page 26: ...nd an 11 pump load Operating time 20 minutes Environmental Operating Transport and Storage Temperature 5 C 40 C 20 C 50 C Humidity 20 90 15 95 Atmospheric Pressure 700 1060hPa 500 1060hPa Classificati...

Page 27: ...13 39 320 Yes 80203UNSxx 235 1240 485 155 14 5 45 5 360 Yes Trolley Compatibility Configuration1 Trolley Compatible 80083UN00 00 5 Alaris Trolley Compatible 80203UNS00 00 6 80203UNSxx 30 80203UNSxx 3...

Page 28: ...unit leakage current must be measured If more than 500 A the equipment should not be used but should be serviced by Qualified Service Personnel Battery It is recommended that only Qualified Service Pe...

Page 29: ...s fluid build up on theWorkstation Do not use aggressive cleaning agents as these may damage the exterior surface of the Workstation Do not steam autoclave ethylene oxide sterilise or immerse this Wor...

Page 30: ...and password are required to access our manuals Please contact a local customer services representative to obtain login details Part Number Description 0000EL00889 AC Power Lead UK 0000EL00938 AC Pow...

Page 31: ...en 8 10 3994 DB Houten Nederland CareFusion 10020 Pacific Mesa Blvd San Diego CA 92121 USA Tel 32 0 2 267 38 99 Tel 34 902 555 660 Tel 31 0 30 2289 711 Tel 1 800 854 7128 Fax 32 0 2 267 99 21 Fax 34 9...

Page 32: ...of CareFusion Corporation or one of its affiliates and its receipt or possession does not convey any rights to reproduce its contents or to manufacture or sell any product described Reproduction discl...

Reviews: