CareFusion Alaris CC User Manual Download Page 31

1000DF00329 Issue 4

30/36

Maintenance (continued)

Before the transfer of the pump to a new patient and periodically during the use, clean the pump by wiping over with a lint-free cloth lightly 

dampened with warm water and a standard disinfectant / detergent solution. 

Recommended cleaners are:
 

Brand  

Concentration

 

Hibiscrub 

20% (v/v)

 Virkon 

 1% 

(w/v)

Do not use the following disinfectant types:

 

• 

Disinfectants which are known to be corrosive to metals must not be used, which include:

  • 

NaDcc (such as Presept), 

  • 

Hypochlorites (such as Chlorasol), 

  • 

Aldehydes (such as Cidex), 

  • 

Cationic Surfactants (such as Benzalkonium Chloride).  

 • 

Use of Iodine (such as Betadine) will cause surface discoloration.  

 • 

Concentrated Isopropyl alcohol based cleaners will degrade plastic parts.

The syringe and extension sets are disposable single use items and should be discarded after use according to their manufacturers’ 

instructions.
If the pump is to be stored for an extended period it should be first cleaned and the internal battery fully charged. Store in a clean, dry 

atmosphere at room temperature and, if available, employ the original packaging for protection.
Once every 3 months during storage, carry out functional tests as described in the Technical Service Manual and ensure that the internal 

battery is fully charged.

A

Before cleaning always switch OFF and disconnect from the AC power supply.  Never allow liquid to enter the 

casing and avoid excess fluid build up on the pump.  Do not use aggressive cleaning agents as these may damage 

the exterior surface of the pump.  Do not steam autoclave, ethylene oxide sterilise or immerse this pump in any 

fluid.
Ensure the pressure transducer is free from residues, which may prevent correct operation of the disc detector.

Cleaning and Storage

Disposal

Information on Disposal for Users of Waste Electrical and Electronic Equipment 

This 

U

 symbol on the product and/or accompanying documents means that used electrical and electronic products should 

not be mixed with household waste. 
If you wish to discard electrical and electronic equipment, please contact your CareFusion affiliate office or distributor for 

further information. 
Disposing of this product correctly will help to save valuable resources and prevent any potential negative effects on human 

health and the environment which could otherwise arise from inappropriate waste handling. 

Information on Disposal in Countries outside the European Union 

This symbol is only valid in the European Union. The product should be disposed of taking environmental factors into 

consideration. To ensure no risk or hazard, remove the internal rechargeable battery and the Nickel Metal Hydride battery 

from the control board and dispose of as outlined by the local country regulations. All other components can be safely 

disposed of as per local regulations. 

Summary of Contents for Alaris CC

Page 1: ...s Alaris CC Syringe Pump Directions For Use en...

Page 2: ...Main Display Features 6 Operating Precautions 7 Getting Started 9 Basic Features 14 Alarms and Warnings 18 Configured Options 19 Specifications 24 Compatible Syringes 25 Associated Products 25 Compati...

Page 3: ...ide 1 Press the a button to turn the pump on 2 CLEAR SETUP NO retains previous data YES clears previous data 3 Load syringe 4 Confirm correct size and brand of syringe 5 Ensure extension set is attach...

Page 4: ...RGE BOLUS MUTE PRESSURE OPTION Extension set hook Pressure transducer Finger Grips Extension set hook Release lever for Rotating Cam Carrying Handle IR Communications port RS232 Connector optional Fol...

Page 5: ...Pump is on hold l Extension set is not connected to the patient l Volume Infused VI is not added BOLUS fluid or drug delivered at an accelerated rate l Pump is infusing l Extension set is connected t...

Page 6: ...ation proof type CF applied part Degree of protection against electrical shock O Protected against vertically falling drops of water r Alternating Current s Device complies with the requirements of Co...

Page 7: ...G DISPLAY icon Indicates time before syringe will require replacing N BATTERY icon Indicates battery charge level to highlight when the battery will require recharging Pump Status Syringe type fitted...

Page 8: ...which may be in the syringe To protect against the introduction of air the user should regularly monitor the progress of the infusion syringe extension line and patient connections and follow the pri...

Page 9: ...its RF emissions are very low and are not likely to cause any interference with the nearby electronic equipment However this pump emits a certain level of electromagnetic radiation which is within the...

Page 10: ...pply for at least 2 hours to ensure that the internal battery is charged verify that the S is lit Language Selection 1 On initial start up the pump will display the Select Language screen 2 Select the...

Page 11: ...ar Rotating cam Release lever push to release Getting Started continued 1 Pull the folded pole clamp towards you and unscrew the clamp to leave enough room for the size of the pole 2 Place pump around...

Page 12: ...accuracy of the infusion rate and may also affect pump performance When drawing fluid into the syringe draw enough to compensate for any dead space volume in the extension set and syringe at the end o...

Page 13: ...original position 7 Ensure that the syringe type and size match those displayed on the pump then press CONFIRM If required the make of syringe can be changed by pressing the TYPE softkey CONFIRM TYPE...

Page 14: ...T THE PRESSURE DISC 5 CONFIRM SYRINGE Check that the syringe type and size being used matches the display If required the make of syringe can be changed by pressing the TYPE button Press CONFIRM when...

Page 15: ...The bolus volume is added to the total volume infused BOLUS Enabled Hands On and Hands Free The Hands Free Bolus is delivered with a single press of the flashing BOLUS soft key On intial use the bolu...

Page 16: ...splays the pressure at a given time 5 Press the OK softkey to exit the pressure screen _ OK 20mmHg 07 45 Pressure Level without pressure set fitted not applicable when FULLY DEDICATED 1 To check and a...

Page 17: ...o confirm the selection 2 Press the YES softkey to select PROTOCOL This will select the pre defined protocol for the selected drug 3 Enter the dosing information prompted on the display for the select...

Page 18: ...softkey to enter the value 5 Select the rate at VTBI end from the list using the f keys and press the OK softkey The default is STOP Set VTBI over Time Event Log This option allows the event log to b...

Page 19: ...battery is exhausted Reconnect to the AC power supply to continue operation and charge the internal battery BATTERY EMPTY The internal battery is exhausted Connect the pump to the AC power supply NEAR...

Page 20: ...n enabled the rate can be locked to prevent unwanted changes of the set infusion rate QUIET MODE When enabled the button beeps are muted AC FAIL When enabled the AC Power Failure Alarm will sound if t...

Page 21: ...AUTO OFFSET 30mmHg 15mmHg 100mmHg PRESSURE DEFAULT 300mmHg 300mmHg 1mmHg 1000mmHg MAX PRESSURE 1000mmHg 1000mmHg 1mmHg 1000mmHg WEIGHT 70Kg 1 00Kg 0 01Kg 250Kg CAP RATE Max infusion rate 1200ml h 1 0...

Page 22: ...may be used at any time to return to the previous screen of the drug library set up procedure Dosing Conversion 1 0 g 1000 ng 1 0 mg h 24 0 mg 24 h 1 0 mg min 60 0 mg h 1 0 mg 1000 g The availability...

Page 23: ...of syringe permitted for use on the pump Select all possible syringes which may be used and disable any that should not be used 1 Select ENABLE SYRINGES from the Configured Options menu using the f ke...

Page 24: ...etup No 1 50 Drug Name 12 Chars max Dose Rate Concentration Bolus Pressure Alarm Dose Units Max Default Min Units Min Default Max Max ml Rate ml h Serial Number Software Version Approved by Configured...

Page 25: ...der normal conditions 95 confidence 95 of pumps Specifications Electrical Classification Class I product Continuous Mode Operation Transportable System Accuracy Volumetric Mean 2 nominal Derating Temp...

Page 26: ...ringe is a specialised syringe with a large diameter barrel To provide protection against accidental dislodging always ensure the infusion line is secured using the infusion set hook see Loading a Syr...

Page 27: ...rd Sets G40015 Standard PVC Syringe Extension Set 150 cm Priming Volume 2 6ml G40020B Standard PVC Syringe Extension Set 200 cm Priming Volume 1 5ml G402EP Extension set Luer lock connectors Kink resi...

Page 28: ...Carefully read the Directions For Use supplied with the extension set prior to use Low Sorbing Sets G40615 Polyethylene Syringe Extension Set 150 cm Priming Volume 1 5ml G40620 Polyethylene Syringe Ex...

Page 29: ...with clamp 152 cm Priming Volume 0 5ml 30832 PVC Y Syringe Extension Set with back check valve and 2 clamps 178 cm Priming Volume 1 5ml 30842E PVC Syringe Extension Set with back check valve SmartSit...

Page 30: ...d damage or premature wear or malfunctions or otherwise operates incorrectly as a result of use with a battery pack that is not manufactured by CareFusion 95 lower confidence interval of 5 hours 50 mi...

Page 31: ...switch OFF and disconnect from the AC power supply Never allow liquid to enter the casing and avoid excess fluid build up on the pump Do not use aggressive cleaning agents as these may damage the ext...

Page 32: ...son the occlusion alarm needs to be set with an operating margin of at least one level between the alarm and normal infusion levels The ability to set a small operating margin permits short time to al...

Page 33: ...erfailureoftheRS232cable Theprotocol is detailed in the Technical Service Manual and is for general information only Any connected analogue and digital componentsarerequiredtomeetIEC EN60950 for data...

Page 34: ...curve As the observation window is reduced short term fluctuations have greater effects as represented by the mouth of the trumpet Knowledge of system accuracy over various observation windows may be...

Page 35: ...hensive list of spare parts for this pump is included within the Technical Service Manual The Technical Service Manual 1000SM00001 is now available in electronic format on the World Wide Web at www ca...

Page 36: ...DB Houten Nederland CareFusion 10020 Pacific Mesa Blvd San Diego CA 92121 USA Tel 32 0 2 267 38 99 Tel 34 902 555 660 Tel 31 0 30 2289 711 Tel 1 800 854 7128 Fax 32 0 2 267 99 21 Fax 34 902 555 661 Fa...

Page 37: ...1000DF00329 Issue 4 36 36 Page Intentionally Left Blank...

Page 38: ...orporation or one of its affiliates and its receipt or possession does not convey any rights to reproduce its contents or to manufacture or sell any product described Reproduction disclosure or use ot...

Reviews: