1000DF00329 Issue 4
17/36
Basic Features (Continued)
If Rate Titration is
enabled
the rate can be adjusted
while infusing
:
1. Select the new rate using the
f
keys.
The message < START TO CONFIRM > will flash on screen and pump continues to infuse at the original rate.
2. Press the
b
button
to confirm the new infusion rate and start infusing at the new rate.
If Rate Titration is
disabled
the rate can only be adjusted
whilst on hold:
1. Press the
h
button to put the pump on hold.
2. Select the new rate using the
f
keys.
3. Press the
b
button
to start infusing at the new rate.
Rate Titration
This option allows a specific VTBI and delivery time to be set. The rate necessary to deliver the required volume within the specified time
is calculated and displayed.
1. Stop the infusion. Press the
d
button to access the options menu.
2. Select the
SET VTBI OVER TIME
option using the
f
keys and press the
OK
softkey.
3. Adjust the volume to be infused using the
f
keys. When the desired volume has been reached press the
OK
softkey.
4. Enter the time over which the volume is to be infused. The infusion rate will automatically be calculated. Press the
OK
softkey to enter
the value.
5. Select the rate at VTBI end from the list using the
f
keys and press the
OK
softkey. The default is STOP.
? Set VTBI over Time
? Event Log
This option allows the event log to be reviewed. It can be enabled/disabled.
1. Press the
d
button to access the options menu.
2. Select the
EVENT LOG
option using the
f
keys and press the
OK
softkey.
3. Scroll through the log using the
f
keys. Press the
QUIT
softkey to exit the log.
? 24 Hour Log
This option allows the 24 hour log of volume infused to be reviewed.
1. Press the
d
button to access the options menu.
2. Select the
24H LOG
option using the
f
keys and press the
OK
softkey.
The display shows the hourly volume infused. The volume infused shown in brackets is the total volume infused since the volume was
last cleared. See example below:
07:48 - 08:00 4.34ml (4.34ml)
08:00 - 09:00 2.10ml (6.44ml)
09:00 - 10:00 2.10ml (8.54ml)
VOLUME CLEARED
3. Press the
QUIT
softkey to exit the log.
? Dosing Summary
To review currently selected dosing information:
1. Press the
d
button to access the options menu.
2. Select
DOSING SUMMARY
.
3. Review the information and then press the
QUIT
softkey.
? Drug Name
To display selected drug name:
1. Press the
d
button to access the options menu.
2. Select
DRUG NAME
.
3. Press the
OK
softkey to confirm the drug name or press the
QUIT
softkey to exit the option.