CareFusion Alaris CC User Manual Download Page 15

1000DF00329 Issue 4

14/36

Basic Features

The 

i

 button allows the delivery of a limited volume of fluid in order to purge the extension set prior to being connected to a patient or 

after changing a syringe.
1.  Press the 

i

 button when the pump is not infusing. Ensure that the extension set is not connected to the patient.

2.  Press and hold the 

PURGE

 softkey until fluid flows and the purging of the IV infusion set is complete. The volume used during purging 

will be displayed, but it is not added to the volume infused.

3.  When purging is complete release the 

PURGE

 softkey. Press the 

QUIT

 softkey to exit back to the main display.

A

The pump will not purge if the rate lock has been enabled.  During PURGE the pressure limit alarms are temporarily 

increased to their maximum level.

Bolus Infusion

Purge

Bolus 

-  Administering a controlled volume of fluid or drug at an increased rate for diagnostic or therapeutic purposes. The pump should 

always be infusing and always attached to the patient. (Drugs given by an IV bolus could achieve immediate and high drug 

concentration levels.)

Bolus can be used at the start of an infusion or during an infusion.

The bolus feature can be configured to:

a) 

BOLUS Disabled

b) 

BOLUS Enabled   i) Hands On only

   ii) 

Hands On and Hands Free

BOLUS Disabled

If configured to 

Disabled

, pressing the 

i

 button will have no effect and the pump will continue to infuse at the set rate.

A

A Hands On bolus and Hands Free bolus cannot be administered if the "RATE LOCK" is active. 
During BOLUS the pressure limit alarm is temporarily increased to the maximum level.

BOLUS Enabled - Hands On

In "Hands on" Bolus, press and hold the (flashing) 

BOLUS

 soft key to deliver the required bolus. The bolus rate can be adjusted. The bolus 

volume is limited in the configuration. 

1.  During infusion press the 

i

 button once to display the bolus screen.

2.  Use the 

f

 keys to adjust the bolus rate if required.

3.  To deliver the bolus press and hold the 

BOLUS

 softkey. During the bolus, the volume being infused is displayed. When the desired bolus 

volume has been delivered or the bolus volume limit is reached, release the softkey.  The bolus volume is added to the total volume 

infused.

BOLUS Enabled - Hands On and Hands Free

The "Hands Free" Bolus is delivered with a single press of the (flashing) 

BOLUS

 soft key.  On intial use the bolus rate and bolus volume are 

at default values and can be changed. On subsequent uses, the bolus rate and bolus volume will remain as per previously set and can be 

changed as required.  Following 

CLEAR SETUP

, the default bolus rate is determined via the configuration and the default bolus volume is 

0.1ml. 

1.  During infusion press the 

i

 button to display the "Hands Free" bolus selection screen.

2.  Press the 

YES

 softkey to go to "Hands Free" selection bolus screen, press the 

HANDS ON

 softkey for "Hands On" bolus (see section 

above).

3.  Use the 

f

 keys to set the bolus volume/dose required; If necessary press the 

RATE

 softkey to adjust the bolus delivery rate 

(150/300/600/900/1200ml/h). 

Note:

 Rate may be restricted by the syringe size and the 

CAP BOLUS RATE

.

4.  Press the flashing 

BOLUS

 softkey once to begin the delivery of the preset bolus. The display will show the bolus being delivered, the 

bolus counting down and  revert to main infusion display upon completion of the bolus. 

5.  To terminate a bolus being delivered press 

STOP

 softkey. This will stop the bolus and continue infusing at the set rate. Press the 

button to stop the bolus delivery and place the pump on hold. 

6.   If the bolus volume reaches the set bolus volume limit the bolus will stop and the pump will revert to infuse at the set infusion rate and 

continue infusing. 

A

If the "Hands Free" bolus option is active, then this feature will be cancelled following any interruption in delivery, 

e.g. occlusion, even if the bolus delivery is incomplete. 
If the volume to be infused (VTBI) is reached during a bolus, the VTBI complete alarm will sound. Press c to silence 

the alarm or CANCEL to acknowledge the alarm. See 'VTBI' section for more details on VTBI operation.

Manual Bolus

The "Manual Bolus" is delivered by moving the plunger drive mechanism forward while the pump is infusing. This method of delivering a 

bolus is not recommended as best clinical practice.

The syringe must be confirmed and the plunger mechanism has to move from an engaged position to disengage and then re-engage 

position. 

A minimum travel of 1mm (leadscrew pitch) must be detected to register.

Summary of Contents for Alaris CC

Page 1: ...s Alaris CC Syringe Pump Directions For Use en...

Page 2: ...Main Display Features 6 Operating Precautions 7 Getting Started 9 Basic Features 14 Alarms and Warnings 18 Configured Options 19 Specifications 24 Compatible Syringes 25 Associated Products 25 Compati...

Page 3: ...ide 1 Press the a button to turn the pump on 2 CLEAR SETUP NO retains previous data YES clears previous data 3 Load syringe 4 Confirm correct size and brand of syringe 5 Ensure extension set is attach...

Page 4: ...RGE BOLUS MUTE PRESSURE OPTION Extension set hook Pressure transducer Finger Grips Extension set hook Release lever for Rotating Cam Carrying Handle IR Communications port RS232 Connector optional Fol...

Page 5: ...Pump is on hold l Extension set is not connected to the patient l Volume Infused VI is not added BOLUS fluid or drug delivered at an accelerated rate l Pump is infusing l Extension set is connected t...

Page 6: ...ation proof type CF applied part Degree of protection against electrical shock O Protected against vertically falling drops of water r Alternating Current s Device complies with the requirements of Co...

Page 7: ...G DISPLAY icon Indicates time before syringe will require replacing N BATTERY icon Indicates battery charge level to highlight when the battery will require recharging Pump Status Syringe type fitted...

Page 8: ...which may be in the syringe To protect against the introduction of air the user should regularly monitor the progress of the infusion syringe extension line and patient connections and follow the pri...

Page 9: ...its RF emissions are very low and are not likely to cause any interference with the nearby electronic equipment However this pump emits a certain level of electromagnetic radiation which is within the...

Page 10: ...pply for at least 2 hours to ensure that the internal battery is charged verify that the S is lit Language Selection 1 On initial start up the pump will display the Select Language screen 2 Select the...

Page 11: ...ar Rotating cam Release lever push to release Getting Started continued 1 Pull the folded pole clamp towards you and unscrew the clamp to leave enough room for the size of the pole 2 Place pump around...

Page 12: ...accuracy of the infusion rate and may also affect pump performance When drawing fluid into the syringe draw enough to compensate for any dead space volume in the extension set and syringe at the end o...

Page 13: ...original position 7 Ensure that the syringe type and size match those displayed on the pump then press CONFIRM If required the make of syringe can be changed by pressing the TYPE softkey CONFIRM TYPE...

Page 14: ...T THE PRESSURE DISC 5 CONFIRM SYRINGE Check that the syringe type and size being used matches the display If required the make of syringe can be changed by pressing the TYPE button Press CONFIRM when...

Page 15: ...The bolus volume is added to the total volume infused BOLUS Enabled Hands On and Hands Free The Hands Free Bolus is delivered with a single press of the flashing BOLUS soft key On intial use the bolu...

Page 16: ...splays the pressure at a given time 5 Press the OK softkey to exit the pressure screen _ OK 20mmHg 07 45 Pressure Level without pressure set fitted not applicable when FULLY DEDICATED 1 To check and a...

Page 17: ...o confirm the selection 2 Press the YES softkey to select PROTOCOL This will select the pre defined protocol for the selected drug 3 Enter the dosing information prompted on the display for the select...

Page 18: ...softkey to enter the value 5 Select the rate at VTBI end from the list using the f keys and press the OK softkey The default is STOP Set VTBI over Time Event Log This option allows the event log to b...

Page 19: ...battery is exhausted Reconnect to the AC power supply to continue operation and charge the internal battery BATTERY EMPTY The internal battery is exhausted Connect the pump to the AC power supply NEAR...

Page 20: ...n enabled the rate can be locked to prevent unwanted changes of the set infusion rate QUIET MODE When enabled the button beeps are muted AC FAIL When enabled the AC Power Failure Alarm will sound if t...

Page 21: ...AUTO OFFSET 30mmHg 15mmHg 100mmHg PRESSURE DEFAULT 300mmHg 300mmHg 1mmHg 1000mmHg MAX PRESSURE 1000mmHg 1000mmHg 1mmHg 1000mmHg WEIGHT 70Kg 1 00Kg 0 01Kg 250Kg CAP RATE Max infusion rate 1200ml h 1 0...

Page 22: ...may be used at any time to return to the previous screen of the drug library set up procedure Dosing Conversion 1 0 g 1000 ng 1 0 mg h 24 0 mg 24 h 1 0 mg min 60 0 mg h 1 0 mg 1000 g The availability...

Page 23: ...of syringe permitted for use on the pump Select all possible syringes which may be used and disable any that should not be used 1 Select ENABLE SYRINGES from the Configured Options menu using the f ke...

Page 24: ...etup No 1 50 Drug Name 12 Chars max Dose Rate Concentration Bolus Pressure Alarm Dose Units Max Default Min Units Min Default Max Max ml Rate ml h Serial Number Software Version Approved by Configured...

Page 25: ...der normal conditions 95 confidence 95 of pumps Specifications Electrical Classification Class I product Continuous Mode Operation Transportable System Accuracy Volumetric Mean 2 nominal Derating Temp...

Page 26: ...ringe is a specialised syringe with a large diameter barrel To provide protection against accidental dislodging always ensure the infusion line is secured using the infusion set hook see Loading a Syr...

Page 27: ...rd Sets G40015 Standard PVC Syringe Extension Set 150 cm Priming Volume 2 6ml G40020B Standard PVC Syringe Extension Set 200 cm Priming Volume 1 5ml G402EP Extension set Luer lock connectors Kink resi...

Page 28: ...Carefully read the Directions For Use supplied with the extension set prior to use Low Sorbing Sets G40615 Polyethylene Syringe Extension Set 150 cm Priming Volume 1 5ml G40620 Polyethylene Syringe Ex...

Page 29: ...with clamp 152 cm Priming Volume 0 5ml 30832 PVC Y Syringe Extension Set with back check valve and 2 clamps 178 cm Priming Volume 1 5ml 30842E PVC Syringe Extension Set with back check valve SmartSit...

Page 30: ...d damage or premature wear or malfunctions or otherwise operates incorrectly as a result of use with a battery pack that is not manufactured by CareFusion 95 lower confidence interval of 5 hours 50 mi...

Page 31: ...switch OFF and disconnect from the AC power supply Never allow liquid to enter the casing and avoid excess fluid build up on the pump Do not use aggressive cleaning agents as these may damage the ext...

Page 32: ...son the occlusion alarm needs to be set with an operating margin of at least one level between the alarm and normal infusion levels The ability to set a small operating margin permits short time to al...

Page 33: ...erfailureoftheRS232cable Theprotocol is detailed in the Technical Service Manual and is for general information only Any connected analogue and digital componentsarerequiredtomeetIEC EN60950 for data...

Page 34: ...curve As the observation window is reduced short term fluctuations have greater effects as represented by the mouth of the trumpet Knowledge of system accuracy over various observation windows may be...

Page 35: ...hensive list of spare parts for this pump is included within the Technical Service Manual The Technical Service Manual 1000SM00001 is now available in electronic format on the World Wide Web at www ca...

Page 36: ...DB Houten Nederland CareFusion 10020 Pacific Mesa Blvd San Diego CA 92121 USA Tel 32 0 2 267 38 99 Tel 34 902 555 660 Tel 31 0 30 2289 711 Tel 1 800 854 7128 Fax 32 0 2 267 99 21 Fax 34 902 555 661 Fa...

Page 37: ...1000DF00329 Issue 4 36 36 Page Intentionally Left Blank...

Page 38: ...orporation or one of its affiliates and its receipt or possession does not convey any rights to reproduce its contents or to manufacture or sell any product described Reproduction disclosure or use ot...

Reviews: