1000DF00329 Issue 4
14/36
Basic Features
The
i
button allows the delivery of a limited volume of fluid in order to purge the extension set prior to being connected to a patient or
after changing a syringe.
1. Press the
i
button when the pump is not infusing. Ensure that the extension set is not connected to the patient.
2. Press and hold the
PURGE
softkey until fluid flows and the purging of the IV infusion set is complete. The volume used during purging
will be displayed, but it is not added to the volume infused.
3. When purging is complete release the
PURGE
softkey. Press the
QUIT
softkey to exit back to the main display.
A
The pump will not purge if the rate lock has been enabled. During PURGE the pressure limit alarms are temporarily
increased to their maximum level.
Bolus Infusion
Purge
Bolus
- Administering a controlled volume of fluid or drug at an increased rate for diagnostic or therapeutic purposes. The pump should
always be infusing and always attached to the patient. (Drugs given by an IV bolus could achieve immediate and high drug
concentration levels.)
Bolus can be used at the start of an infusion or during an infusion.
The bolus feature can be configured to:
a)
BOLUS Disabled
b)
BOLUS Enabled i) Hands On only
ii)
Hands On and Hands Free
BOLUS Disabled
If configured to
Disabled
, pressing the
i
button will have no effect and the pump will continue to infuse at the set rate.
A
A Hands On bolus and Hands Free bolus cannot be administered if the "RATE LOCK" is active.
During BOLUS the pressure limit alarm is temporarily increased to the maximum level.
BOLUS Enabled - Hands On
In "Hands on" Bolus, press and hold the (flashing)
BOLUS
soft key to deliver the required bolus. The bolus rate can be adjusted. The bolus
volume is limited in the configuration.
1. During infusion press the
i
button once to display the bolus screen.
2. Use the
f
keys to adjust the bolus rate if required.
3. To deliver the bolus press and hold the
BOLUS
softkey. During the bolus, the volume being infused is displayed. When the desired bolus
volume has been delivered or the bolus volume limit is reached, release the softkey. The bolus volume is added to the total volume
infused.
BOLUS Enabled - Hands On and Hands Free
The "Hands Free" Bolus is delivered with a single press of the (flashing)
BOLUS
soft key. On intial use the bolus rate and bolus volume are
at default values and can be changed. On subsequent uses, the bolus rate and bolus volume will remain as per previously set and can be
changed as required. Following
CLEAR SETUP
, the default bolus rate is determined via the configuration and the default bolus volume is
0.1ml.
1. During infusion press the
i
button to display the "Hands Free" bolus selection screen.
2. Press the
YES
softkey to go to "Hands Free" selection bolus screen, press the
HANDS ON
softkey for "Hands On" bolus (see section
above).
3. Use the
f
keys to set the bolus volume/dose required; If necessary press the
RATE
softkey to adjust the bolus delivery rate
(150/300/600/900/1200ml/h).
Note:
Rate may be restricted by the syringe size and the
CAP BOLUS RATE
.
4. Press the flashing
BOLUS
softkey once to begin the delivery of the preset bolus. The display will show the bolus being delivered, the
bolus counting down and revert to main infusion display upon completion of the bolus.
5. To terminate a bolus being delivered press
STOP
softkey. This will stop the bolus and continue infusing at the set rate. Press the
h
button to stop the bolus delivery and place the pump on hold.
6. If the bolus volume reaches the set bolus volume limit the bolus will stop and the pump will revert to infuse at the set infusion rate and
continue infusing.
A
If the "Hands Free" bolus option is active, then this feature will be cancelled following any interruption in delivery,
e.g. occlusion, even if the bolus delivery is incomplete.
If the volume to be infused (VTBI) is reached during a bolus, the VTBI complete alarm will sound. Press c to silence
the alarm or CANCEL to acknowledge the alarm. See 'VTBI' section for more details on VTBI operation.
Manual Bolus
The "Manual Bolus" is delivered by moving the plunger drive mechanism forward while the pump is infusing. This method of delivering a
bolus is not recommended as best clinical practice.
The syringe must be confirmed and the plunger mechanism has to move from an engaged position to disengage and then re-engage
position.
A minimum travel of 1mm (leadscrew pitch) must be detected to register.