6-2
Troubleshooting
Alaris
®
Auto-ID Module
Technical Service Manual
6.2
General TrOuBleShOOTInG
(Continued)
• Data Set must have been wirelessly
downloaded to PC Unit, and have proper
aliases and/or NDCs set up for drugs.
Once this is accomplished, scanner
light comes on when a pumping module
channel is accessed to set up a drug.
nOTe:
To help determine whether or not a
problem is related to the Data Set, it would
be helpful to track (if possible) when a new
Data Set is downloaded.
• A Pump Module, Syringe Module, or
PCA Module must be attached as part of
Alaris
®
System.
• PC Unit, pumping module, Auto-ID
Module, and handheld scanner must be
properly and securely connected.
Table 6-1 Software Compatibility
The software versions identified in this table were current at the time this manual was written and may
change over time. Check for Service Bulletins addressing software updates or contact CareFusion
Technical Support if there are any questions.
nOTe:
If the ready indicator does not light for approximately 1 second when the module is first attached, the
Auto-ID Module software is not compatible.
Software Version
(or later)
Model 8000
8.5.25.0
Model 8015
8.15.16.0
Model 8000
9.0.23.0
Model 8015
9.0.23.0
Auto-ID Module
8.5.26.0
X
X
X
X
Wireless Main
8.14.10.0
9.0.20.0
X
Not Applicable
X
Not Applicable