![CareFusion Alaris 8210 Series Technical & Service Manual Download Page 5](http://html1.mh-extra.com/html/carefusion/alaris-8210-series/alaris-8210-series_technical-and-service-manual_2557167005.webp)
1-1
Alaris
®
SpO
2
Modules
Technical Service Manual
Chapter 1 – GENERAL INFORMATION
1.1 INTRODUCTION
This manual describes how to service the
S
p
O
2
Modules (
8210
and
8220
Series). It is
used in conjunction with the following Alaris
®
System documents and software:
• PC Unit / Pump Module (
8000
/
8100
Series) Technical Service Manual
• Alaris
®
System Directions for Use (
DFU
)
• Maintenance Software and User Manual
(v
8.1
or later)
This manual is intended for personnel
experienced in analysis, troubleshooting
and repair of analog/digital microprocessor-
based electronic equipment.
Reference the Alaris
®
System
DFU
for a
product introduction, detailed setup and
operation procedures, defi nitions (including
precaution defi nitions), specifi cations, and
other information related to the use of the
Alaris
®
System.
If the
S
p
O
2
Module requires service while
under warranty, it must be serviced only by
CareFusion authorized service personnel.
Reference the "Warranty" and "Service
Information" sections in the Alaris
®
System
DFU
.
1.2 FEATURES
• Functional blood oxygen saturation
percentage, pulse rate, pulse bar,
channel identification and scrolling
messages display (such as, alerts and
alarms).
• Alarm limit settings, trend data and
plethysmographic waveforms display.
•
8210 Series:
SatSeconds alarm
Management Technology, to allow
monitoring of
%S
p
O
2
below selected low
alarm limit for a period of time before an
audible alarm sounds.
CAUTION
To
avoid damaging the keypad
, do not use
sharp objects (such as, pens, pencils) to activate
switches.
CAUTION
Any attempt to service this product by anyone
other than an authorized CareFusion Service
Representative, while the product is under
warranty,
may invalidate the warranty
.
Summary of Contents for Alaris 8210 Series