2-4
Alaris
®
SpO
2
Modules
Technical Service Manual
CHECKOUT AND CONFIGURATION
System Configuration - SPO2
>Press CONFIRM
Louder
Pulse Beep
Volume
Softer
CONFIRM
Off
4. To accept change, press
ENTER
key.
• Display highlights next limit and
prompts for an entry.
5. To change another alarm limit, press
soft key for that parameter and make
necessary change.
6. To accept change(s), press
CONFIRM
soft key.
2.4.3 Limit
Mode
1. After
accessing
System Config -
SPO2
options display, press
Limits Mode
soft
key.
• Limit Mode cannot be changed
if a Profi le is being used for
programming.
2. Select
Adult
or
Neonatal
by pressing
applicable soft key.
• Following display represents
Neonatal alarm limits
but same
procedure is used for confi guring
Adult alarm limits.
3. To change an alarm limit, see "Alarm
Limits" section.
OR
To accept settings, press
CONFIRM
soft
key.
2.4.4 Pulse Beep Volume
1. After
accessing
System Config -
SPO2
options display, press
Pulse beep
volume
soft key
• Following display refl ects that
Pulse
Beep Volume
is
Off
. To display
volume options, press
Louder
soft
key.
2.4 CONFIGURATION
SETUP
(Continued)
2.4.2 Alarm
Limits
(Continued)
System Configuration - SPO2
PULSE
LOW
>Enter High Pulse Limit
98
90
150
50
%SPO2
HIGH
%SPO2
LOW
PULSE
HIGH
Adult alarm limits
CONFIRM
System Configuration - SPO2
Allows selection
between the
adult and neonatal
modes
of operation.
Adult
Neonatal
Limit Mode Setup
System Configuration - SPO2
Neonatal alarm limits
%SPO2
HIGH
%SPO2
LOW
88
70
120
40
PULSE
HIGH
PULSE
LOW
>Select Parameter Limit
CONFIRM